Command Reference
R&S
®
ZNA
1131
User Manual 1178.6462.02 ─ 12
Note:
The receiver calibration relies on the source power calibration acquired using
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Port number.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Event
Manual operation:
See
"Harmonic Orders to Calibrate"
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note:
The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
is enabled
only after steps 1 and 2 have been carried out (
CORRection:IMODulation:LTONe[:ACQuire]
CORRection:IMODulation:UTONe[:ACQuire]
).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
).
Setting parameters:
<CalOnlyPortFreq>
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNA-K4).
SCPI Command Reference