Command Reference
R&S
®
ZNA
1104
User Manual 1178.6462.02 ─ 12
Suffix:
<Ch>
.
Channel number
Setting parameters:
<CalKitName>
The name of the cal kit to be loaded
<KitLabel>
The label of the cal kit to be loaded
<Gender>
MALE | FEMale | NGENder
The gender of the cal kit to be loaded
<Port>
Either a comma-separated list of port numbers or ALL (optional)
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:CKIT:PORT<PhyPt>?
[<Detail>]
Queries the cal kit data assigned to the given port via
●
[SENSe<Ch>:]CORRection:COLLect:CKIT:LOAD
[SENSe<Ch>:]CORRection:COLLect:CKIT:INSTall
Suffix:
<Ch>
.
Channel number
<PhyPt>
Port number
Query parameters:
<Detail>
CONNector | LABel | NAME | GENDer
Queried property
Return values:
<Result>
Usage:
Query only
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
<ConnectorType>
Selects a connector type at a specified port
<PhyPt>
and its gender.
Tip:
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
to select
an arbitrary connector type using a string variable.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Port numbers of the analyzer
Note:
If the analyzer is set to use the same connectors at all
[SENSe<Ch>:]CORRection:COLLect:CONNection:
ALL
), then a change of a connector type is valid for all
ports. The gender of the connectors can still be different.
SCPI Command Reference