Command Reference
R&S
®
ZNA
1029
User Manual 1178.6462.02 ─ 12
Setting parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a network analyzer-specific cal kit
file with the extension
*.calkit
.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New Kit
1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
... :MMEM:STOR:CKIT 'New Kit 1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New Kit
1.calkit'
Store the data for the user-defined cal kit "New Kit 1" and over-
write the cal kit file
New Kit 1.calkit
.
Usage:
Setting only
Manual operation:
See
"Import Cal Kit... / Export Cal Kit..."
MMEMory:STORe:CKIT:WLABel
<CalKitName>, <KitLabel>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by
its name and label.
Setting parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<KitLabel>
Label of the calibration kit, usually its serial number.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a NWA-specific cal kit file with the
extension *.calkit.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
Example:
See
[SENSe:]CORRection:CKIT:LLABel
Usage:
Setting only
MMEMory:STORe:CMAP
<ColorSchemeFile>
Stores a color scheme to a specified VNA color scheme file.
SCPI Command Reference