Remote Commands
R&S
®
Spectrum Rider FPH
375
User Manual 1321.1011.02 ─ 11
Example:
POW:ACH:PRES:RLEV
Initiates an ACP measurement and adjusts the reference level.
Usage:
Event
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
<TxReferenceMode>
This command automatically selects a reference channel for relative ACLR measure-
ments.
The number of transmission channels and adjacent
/
alternate channels has to be
greater than 0 for the command to work.
Parameters:
<TxReferenceMode> MANual | MINimum | MAXimum | LHIGhest
MANual
Defines a reference channel for relative ACLR measurements.
You need at least one channel for the command to work.
MINimum
Reference is the transmission channel with the lowest power.
MAXimum
Reference is the transmission channel with the highest power.
LHIGhest
Reference is the lowest transmission channel for lower adjacent
channels and highest transmission channel for upper adjacent
channels.
Example:
POW:ACH:REF:TXCH:AUTO MIN
Sets the TX channel with the lowest power as reference chan-
nel.
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
<TxReferenceChannel>
This command defines a reference channel for relative ACLR measurements.
The number of transmission channels and adjacent
/
alternate channels has to be
greater than 0 for the command to work.
Parameters:
<TxRefCh>
The actual value range depends on the number of active trans-
mission channels.
Range:
1 to 12
*RST:
1
Example:
See
[SENSe:]POWer:ACHannel:TXCHannel:COUNt
Remote Control - Commands