Instrument Tour
R&S
®
FSW
46
Getting Started 1338.4102.02 ─ 08
5.1.10
NOISE SOURCE CONTROL
The noise source control female connector is used to provide the supply voltage
for an external noise source. For example, use it to measure the noise figure and
gain of amplifiers and frequency converting devices.
Conventional noise sources require a voltage of +28
V to be switched on and 0
V
to be switched off. The output supports a maximum load of 100
mA.
5.1.11
Probe
The R&S
FSW provides a connector for supply voltages of +15
V to -12
V and
ground for active probes and preamplifiers. A maximum current of 140
mA is
available. This connector is suitable as a power supply for high-impedance
probes.
For details on configuring and using power sensors, see the R&S
FSW User
Manual.
5.1.12
POWER SENSOR
The LEMOSA female connector is used to connect Rohde
&
Schwarz power sen-
sors. For a detailed list of supported sensors, see the data sheet.
For details on configuring and using power sensors, see the User Manual.
5.1.13
USB
The front panel provides three female USB connectors (USB-A) to connect devi-
ces like a keyboard or a mouse. In addition, a memory stick can be connected to
store and reload instrument settings and measurement data.
The rear panel provides further USB connectors, including a male (USB-B)
connector. See
All USB connectors support standard 2.0.
Front Panel View