Preparing for Use
R&S
®
FSW
21
Getting Started 1338.4102.02 ─ 08
The R&S
FSW boots.
After booting, the instrument is ready for operation.
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended warm-
up time (see data sheet).
To shut down the product
The product is in the ready state.
► Press the Power key.
The operating system shuts down. The LED changes to orange.
If the instrument temperature exceeds the limit specified in the data sheet,
the R&S
FSW automatically shuts down to protect the instrument from dam-
age.
To disconnect from power
The R&S
FSW is in the standby state.
1.
NOTICE!
Risk of data loss. If you disconnect the product from power when it
is in the ready state, you can lose settings and data. Shut it down first.
Set the switch on the power supply to position [0].
See
Chapter 5.2.2, "AC Power Supply Connection and Main Power Switch"
on page 50.
The LED of the Power key is switched off.
2. Disconnect the R&S
FSW from the power source.
4.7
Connecting to LAN
You can connect the instrument to a LAN for remote operation via a PC. For
details on the connector, see
Connecting to LAN