Command Reference
R&S
®
ZNB/ZNBT
1147
User Manual 1173.9163.02 ─ 53
[SENSe<Ch>:]FREQuency:CONVersion
<ConversionMode>
Enables a frequency conversion measurement mode for channel
<Ch>
.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConversionMode>
FUNDamental | ARBitrary | MIXer | IMODulation
FUNDamental
Measurement of the fundamental signal (non frequency-convert-
ing mode). This selection cancels all port-specific frequency and
power settings (see example).
ARBitrary
Frequency conversion mode (arbitrary port frequencies). This
mode is automatically activated when a port-specific frequency
is defined.
MIXer
Scalar mixer mode. The mixer mode is configured using the
[SENSe<Ch>:]FREQuency:CONVersion:MIXer...
com-
mands.
IMODulation
Intermodulation measurement mode, configured using the
[SENSe<Ch>:]FREQuency:IMODulation...
commands.
*RST:
FUNDamental
Example:
See
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
Manual operation:
See
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary
<Numerator>, <Denominator>,
<Offset>, <SweepType>
Unused command.
The R&S
ZNB/ZNBT doesn't allow to set a separate receiver frequency for
. Use
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
to define the port-specific frequency (range or CW).
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary:PMETer<Pmtr>
<Numerator>,
<Denominator>, <Offset>, <SweepType>
Defines the receiver frequency of a power meter used for frequency-converting mea-
surements. The receiver frequency is either a range (for frequency sweeps) or a CW
frequency (for power, time and CW Mode sweeps). The receiver frequency is valid for
all ports.
<Numerator>, <Denominator> and <Offset> are parameters of the frequency formula.
The receiver frequency f
r
is calculated according to
SCPI Command Reference