Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.6
E-4
SENSe (Sensor Configuration)
The sensor is configured by means of the commands of the groups
SENSe
and
TRIGger
.
Table 6-2
Commands of the
SENSe
system
Command
Parameter
Unit
Remarks
SENSe
:AVERage
:COUNt[?]
1 to 65536
:AUTO[?]
OFF | ON | ONCE
:MTIMe[?]
0.01 to 999.99
s
:NSRatio[?]
0.0 to 1.0
dB
:RESolution[?]
1 to 4
:SLOT[?]
1 to <SENSe:POWer:TSLot:AVG:COUNt>
:TYPE[?]
RESolution | NSRatio
:RESet
No query
:STATe[?]
OFF | ON
:TCONtrol[?]
MOVing | REPeat
:CORRection
:DCYCle[?]
0.001 to 99.999
%
:STATe[?]
OFF to ON
:OFFSet[?]
-200.0 to 200.0
dB
:STATe[?]
OFF to ON
:SPDevice:STATe[?]
OFF to ON
:FREQuency[?]
10.0e6 to 18.0e9 (NRP-Z28) or
9.0e3 to 6.0e9 (NRP-Z98)
Hz
:FUNCtion[?]
“POWer:AVG“ | “POWer:TSLot:AVG“ |
“POWer:BURSt:AVG“ | “XTIME:POWer“(NRP-Z28)
or
“POWer:AVG“ (NRP-Z98)
:POWer
:AVG
:APERture[?]
10.0e-6 to 0.3
s
:BUFFer