Roche Diagnostics
B-8
Operator’s Manual · Version 1.0
2 Hardware
URISYS 2400
Installation
Liquid waste tank
The liquid waste tank is located outside the analyzer. The capacity of the tank
depends on the specific tank that is used. The tank must have a volume of at
least 5 L (5.3 qts). An overflow sensor is available as an option.
A 5 m (16.4 ft) liquid waste tube is supplied for the connection of the liquid
waste tank to the analyzer.
a
To set up the liquid waste tank
1
Ensure the liquid waste tank is empty before use. The liquid waste tank
must be positioned well below the level of the analyzer, to allow for the
reliable flow of the liquid waste into the tank.
2
Connect the liquid waste tube to the drain socket on the analyzer. If
necessary, shorten the liquid waste tube to ensure smooth and even flow.
3
If an overflow sensor is used, connect it to the drain level connector on the
analyzer and place it into the liquid waste tank.
TS transfer base
a
To set up the TS transfer base and the strip waste container
1
Open the strip waste container door on the right side of the analyzer.
2
Insert the TS transfer base smoothly to its end position, and ensure that it
is positioned correctly.
3
Turn the screw on the TS transfer base
until it locks to secure the base.
4
Insert a waste box carton into the strip waste container and place it into its
correct position.
Figure B-4 Inserting the TS transfer base into the
URISYS 2400
analyzer
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...