Roche Diagnostics
E-12
Operator’s Manual · Version 1.0
9 Instrument alarms (messages)
URISYS 2400
Instrument alarms (messages)
Instrument alarms (messages)
This list shows the messages that can be acted on by the operator. If the screen
shows an alarm code that is not mentioned in this list, contact Technical
Support.
Alarms
Category
Alarm name
Type of error
Subcategory
Level
Check and countermeasure
1
S.Probe
Sample Probe Error
1, 3, 5
S.STOP
1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
2, 4, 6, 8-9, 12-
13
STOP
1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
7
S.STOP
1. Clean the tip of the sample probe.
2. Check the sample probe and the
electrode for damage, or whether they
are in contact with each other.
3. If the condition cannot be restored to
normal, contact Technical Support.
11
STOP
1. Check whether the liquid waste tank is
full, or the waste flow path is blocked.
2
TS Cassette
Test Strip Cassette
Mechanism Error
1-7
STOP
1. Confirm whether a test strip is
misplaced in the cassette mechanism.
Remove the test strip as required.
2. Execute the reset procedure.
3. If the condition cannot be restored to
normal, contact Technical Support.
8-9
WARNING
1. Close the URISYS 2400 cassette
compartment door.
2. If the condition cannot be restored to
normal, contact Technical Support.
10-11
WARNING
1. Insert a URISYS 2400 cassette, shift it
to the rear and execute Initial or Test on
the [TS Cassette] screen.
12-14
STOP
1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
Table E-3
Instrument Alarms (messages)
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...