8 Data alarms (flags)
URISYS 2400
Flags on result data printout
Flags on result data printout
The result data that are printed must be checked for flags. There can be
different flags next to the data, in order to bring some of the values to the
attention of the operator. Flags on result data printout are outlined below.
Flag
Data alarm
Explanation
Appearance
on printout
S
Sieve
Test parameter result corresponds
to sieve criteria settings
After the value
*
Abnormal
Test parameter result corresponds
to abnormal criteria settings
Before the
value
O
Specific gravity
(SG) out of range
Result is greater than 1.050 or less
than 1.000
Instead of SG
value
I
SG value invalid
Refraction position is not definite
Instead of SG
value
T
SG error
Hardware error
Instead of SG
value
#
Changed range
value
Initial range table is changed
After the value
!
Edit
Data have been edited
After the value
Table E-1
Flags on result data printout
Flag
Sample alarm
Explanation
Appearance
on printout
E
Sample empty
No sample detected
Instead of
value
N
Sample short
Sample volume below limit
Instead of
value
R
Test strip
Test strip error
Instead of
value
Table E-2
Sample alarm flags
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...