Maintenance and Trouble Shooting
85
C
C
3.2
UV Decontamination
After cleaning the Reagent/Sample Stage with bleach and water/ethanol, sterilize the
instrument with UV light, as follows:
䉴
Close the instrument door and touch the
UV Decontamination button on the
MAINTENANCE Screen.
䉴
On the DECONTAMINATION SETUP
Screen that appears, use the Up and
Down buttons to set the decontamination
time. The recommended setting is 6 hours.
Note:
In the Time field, one unit is 15 min-
utes.
䉴
Touch the Start button.
䉴
The screen will display the time
remaining in the decontamination
procedure.
Note:
You can stop the decontamination
procedure at any time by touching the
Cancel button.
䉴
After the decontamination procedure is
finished, either return to the MAIN
MENU Screen to perform a purification run
(e.g. when the decontamination has been run overnight) or shut the instrument down.
For safety reasons it is recommended to decontaminate the surface of the workstation
and the accessories additionally by wiping it with decontaminating reagents that destroy
any nucleic acids or DNases/RNases. Also decontaminate the work area where the manual
handling steps are performed.
For decontamination either bleach or the following commercially available reagents may
be used (follow the working procedures given by the respective manufacturer): LTK-
008
TM
(Biodelta GmbH, Germany; www.biodelta.de) DNAZap
TM
or RNase Zap
TM
(Ambion Inc., Texas, USA; www.ambion.com).
Cleaning Instructions and UV Decontamination
UV Decontamination
MagNA Pure Compact Manual.book Seite 85 Dienstag, 28. Februar 2006 4:16 16
Содержание MagNA Pure Compact Instrument
Страница 2: ......
Страница 6: ......
Страница 11: ......
Страница 35: ...www roche applied science com MagNA Pure Compact Operator s Manual Instrument Version 1 0 Software Version 1 1...
Страница 36: ......
Страница 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Страница 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Страница 95: ...Maintenance and Trouble Shooting MagNA Pure Compact Manual book Seite 69 Dienstag 28 Februar 2006 4 16 16...
Страница 129: ......
Страница 130: ...www roche applied science com 04555791001 0306 Roche Diagnostics GmbH Roche Applied Science 68298 Mannheim Germany...