2.1
Installing or replacing the battery pack
When shipped, the battery pack is not installed in the
CoaguChek Pro II meter.
Unused battery packs lose their charge over time and
have to be recharged before they can be used. After
installing a new battery pack, the meter should be
charged for two hours before testing. Please note that
the battery pack will only reach full capacity once it has
been fully emptied and charged several times.
Whenever the meter is placed on an active Handheld
Base Unit or powered by the power adapter, the
icon is displayed. This icon shows that power is availa-
ble and the meter can be charged if necessary.
Make sure that the permitted temperature range for
charging the battery pack (12-32 °C or 54-90 °F) is
maintained during installation and initial setup.
Содержание CoaguCheck Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual ...
Страница 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Страница 6: ...6 This page intentionally left blank ...
Страница 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Страница 42: ...42 This page intentionally left blank ...
Страница 116: ...116 This page intentionally left blank ...
Страница 138: ...138 This page intentionally left blank ...
Страница 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Страница 170: ...170 This page intentionally left blank ...
Страница 174: ...174 This page intentionally left blank ...
Страница 182: ...182 This page intentionally left blank ...
Страница 186: ...186 This page intentionally left blank ...
Страница 187: ......