Roche Diagnostics
E-78
Operator’s Manual · Version 3.1
15 ISE troubleshooting
Cedex Bio System
List of ISE flags
Seg Isecal
e
.
Sol1 F Dev
Meaning
The ISE Solution 1 factor was not within the expected range.
Message ID
22
Priority
18
Possible cause
This flag occurs with calibrations because:
o
The dosing from the sample probe is not accurate.
o
The ISE tower is blocked or the mixing was not sufficient.
o
There may be problems with the ISE unit tubing, the electrodes, or the ISE unit
was not serviced properly.
o
The ISE Calibrator Indirect has expired.
o
ISE Solution 1 has expired.
Recommended actions
If only one calibration is flagged:
o
Repeat the calibration.
If successive calibrations are flagged:
o
Check whether the ISE Calibrator Indirect is in the correct position.
o
Check whether the ISE Calibrator Indirect has expired.
o
Check whether the ISE Solution 1 has expired.
o
Perform the maintenance action
Clean ISE Tower Automatically
.
o
Perform the maintenance action
Activate Electrodes
.
o
Perform the maintenance action
Deproteinize Probe
.
o
Perform the maintenance action
Prime ISE Calibrator & Reference
.
o
Check whether the electrode expiration date has passed. Replace the expired
electrode, if necessary, and perform the maintenance action
Electrode Service
.
o
Check the ISE unit tubing for leaks and blockages.
o
Contact your service representative.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...