Roche Diagnostics
E-74
Operator’s Manual · Version 3.1
15 ISE troubleshooting
Cedex Bio System
List of ISE flags
No Fluid
Meaning
The ISE sample or ISE Reference Solution sensor could not detect any fluid. It takes
too long for a sample segment to reach the electrodes.
Message ID
68
Priority
2
Possible cause
o
The fluid was not pipetted.
o
The ISE tower is clogged.
o
The ISE tubing is blocked or leaking.
o
The ISE sample sensor is defective.
o
The ISE Electrodes are blocked or not properly mounted.
Recommended actions
1. Repeat the measurement.
2. Check the sample tubing for clots.
3. Check ISE Electrodes and O-Rings for proper mounting.
4. Check the ISE tower and ISE tubing for leaks and blockages.
5. Ensure that the ISE bottles on the ISE unit are available and check the level of
fluid; replace if necessary. (Allow at least 15 seconds for priming the ISE bottle
before continuing with any other activity, e.g. replacing the other ISE bottle.)
No Isecal
Meaning
The ISE sample sensor could not detect any calibrator. It takes too long for a
calibrator segment to reach the electrodes.
Message ID
90
Priority
3
Possible cause
o
ISE Calibrator Indirect bottle is not available.
o
The ISE tubing is leaking or blocked.
o
The ISE sample sensor is defect.
Recommended actions
1. Repeat the measurement.
2. Ensure that the ISE Calibrator Indirect on the ISE unit is available and check the
level of fluid; replace if necessary
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...