14 ISE maintenance
ISE maintenance actions
Replace ISE unit tubing
f
Utilities
>
Maintenance
>
Replace ISE Unit Tubing
Due to the abrasion by the ISE valve action, the ISE tubing needs to be replaced
periodically. Replacing the ISE tubing ensures the proper functioning of the ISE unit.
During the
Replace ISE Unit Tubing
maintenance action, the following
maintenance actions are automatically performed:
o
Prime ISE Reference and Calib.
o
Initialize ISE Reference Sensor
o
Condition ISE Tubing
o
Initialize ISE Unit
Operator time
Approximately 15 minutes.
Prerequisites
The system must be in
Standby
status.
Tools and materials required
m
Activator
m
Roche approved ISE unit tubing set
m
Tissues
a
To remove the ISE unit tubing
1
Select the maintenance action
Replace ISE Unit Tubing
.
2
Press .
The maintenance definition screen is displayed.
3
Press
to start the action.
The placement screen is displayed.
4
Place the Activator on the sample area position indicated on the screen.
5
Press
to confirm the placement.
The tubing is emptied.
Your are asked to open the covers.
6
Open the main cover, then the left service flap and the ISE cover.
Your are asked to remove the tubing adapters from the ISE solution bottles.
Always replace the ISE tubing with the help of the
Replace ISE Unit Tubing
maintenance
action.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Infection by biohazardous materials
o
o
Personal injury or damage to the analyzer due to contact with instrument mechanism
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...