Roche Diagnostics
E-12
Operator’s Manual · Version 3.1
12 ISE description
Cedex Bio System
Basic operation
Basic operation
Operation of the ISE unit is integrated in the operation of the main instrument.
The following sections contain some ISE specific information on operation-related
issues.
e
For general information on operating the main instrument
, see
the corresponding
Introduction to the instrument
.
For specific ISE tasks, see Chapter 13
.
Samples
The sample tubes are placed on the sample area of the main instrument.
Calibration
The ISE unit requires frequent calibration to ensure the accuracy of the test results
obtained.
Main calibration
The electrodes are calibrated with a two-point calibration using ISE Solution 1 and 2.
The main calibration is typically performed during the Prepare phase.
One-point calibration
The electrodes are calibrated after each ISE measurement using the on-board ISE
Calibrator indirect/urine. This calibration is an integral part of each ISE
measurement and is performed automatically.
ISE Standby
If there is no measurement for more than three minutes, the ISE unit switches to
Standby
status. In this status, ISE Calibrator indirect/urine and ISE Reference
Solution are pumped into the measuring channel and moved a short distance at
regular intervals. Also, every two hours the ISE tubing is primed with ISE Calibrator
indirect/urine and ISE Reference Solution. This is done to prevent the following
problems:
o
Flow of ISE Reference Solution backwards into the measuring channel, which can
damage the electrodes because of the high ion concentration of the ISE Reference
Solution.
o
Crystallization of salts in the tubing, causing blockages.
The ISE unit has a separate power supply. Therefore, the
Standby
status is
maintained even if the main instrument is switched off.
NOTICE
Damage to electrodes and possible tubing blockage
o
Do not unplug or switch off the ISE power supply. Periodic flow of solutions must be
performed at all times.
o
If you intend not to use the ISE unit for more than one week, you should deactivate it.
This saves ISE fluids, reduces wear and tear of the tubing, and lowers the likelihood of
malfunctions and errors occurring. (See
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...