background image

Roche Diagnostics

 

Operator’s Manual · Version 3.1

E-7

Cedex Bio System

12 ISE description

Overview

Principles of operation

The ISE unit uses flow-through ion-selective electrodes and a reference electrode 
with an open liquid junction. Each electrode has a membrane that is sensitive to a 
particular type of ion.

Measuring process

1. The ISE maintenance and standard solutions (ISE Deproteinizer, ISE Etcher, 

Activator, and ISE Solution 1 and 2) are pipetted from sample tubes on the 
sample area to the ISE tower as required.

2. The sample is pipetted from the sample tube (located on the sample area) into the 

ISE tower. The sample is diluted with system water. Mixing is performed with 
four air jets arranged in a circle. These jets blow air into the tower to produce a 
homogenous mixture.

3. The sample is divided into segments with the aid of a special arrangement of 

valves. The first (shorter) segments are used for cleaning, these are followed by a 
longer segment, on which the measurements are made.

4. The sample is passed to the ion-selective electrodes by the action of the peristaltic 

pump.

The exact positioning of segments is ensured by the ISE sample sensor.

5. In the meantime, the ISE tower is washed with distilled water and dried.

6.

ISE Reference Solution

 is passed through the ISE Reference Electrode and into the 

measuring channel downstream of the electrodes. The 

ISE Reference Solution

 

completes the electrical circuits for each electrode so that measurements can be 
made. While the measurements are made, the sample and 

ISE Reference Solution

 

are stationary.

7. A one-point calibration is performed after each sample measurement using the 

ISE Calibrator Indirect, which is located on ISE unit.

8. The electrolyte concentration of the sample is calculated.

A

ISE Reference Electrode

B

ISE Reference Solution sensor

C

ISE tower

D

Samples, controls, ISE Solution 1 and 2, 
Activator, ISE Etcher, and ISE Deproteinizer

E

ISE Reference Solution

F

ISE Calibrator Indirect

G

Peristaltic pump

H

Measuring channel

I

Ion-selective electrodes

J

Electrode block

K

ISE sample sensor

Figure E-1

Main parts of the ISE measuring system

K

Ref Na

B

G

C

I

D

E

F

K

A

J

H

Содержание Cedex Bio

Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...

Страница 2: ...d Software updates are done by service representatives Intended use The Cedex Bio instrument is a continuous random access analyzer intended for the determination of chemical and biochemical parameter...

Страница 3: ...ng the shift B 10 Preparing the system B 12 Analyzing samples B 34 Validating sample results B 51 Performing calibrations B 59 Performing QC B 68 Finishing the shift B 77 Logging off B 88 Shutting dow...

Страница 4: ...ription Overview E 5 Hardware E 9 Basic operation E 12 Technical specifications E 13 13 ISE operation Daily operation E 17 Replacing ISE fluid bottles E 30 Replacing electrodes E 32 Cleaning the ISE t...

Страница 5: ...nts at the beginning of the manual and each chapter In addition a complete index can be found at the end Online Help system The Cedex Bio instrument has a context sensitive online Help feature to aid...

Страница 6: ...the buttons is used without color or navigation indicators Screenshots The screen representations shown in this publication are for illustrative purposes only The screens do not necessarily show valid...

Страница 7: ...i gratia for example EMC Electromagnetic compatibility EN European standard i e Id est that is to say IEC International Electrical Commission ISE Ion selective electrode LED Light emitting diode LIS L...

Страница 8: ...centigrade L microliter m micrometer A amp re cm centimeter h hour Hz hertz LB pound weight in inch kg kilogram kVA kilo volt ampere L liter m meter MB megabytes min minute mL milliliter mm millimeter...

Страница 9: ...1 Safety A 3 2 Introduction to the instrument A 15 3 Hardware A 37 4 Software A 67 System Description A...

Страница 10: ......

Страница 11: ...peration and maintenance A 6 Biohazardous materials A 6 Waste A 7 Reagents and other working solutions A 7 Installation A 7 Environmental conditions A 7 Power interruption A 8 Electromagnetic devices...

Страница 12: ...Roche Diagnostics A 4 Operator s Manual Version 3 1 1 Safety Cedex Bio System Table of contents Legal liability A 12 Safety labels A 13...

Страница 13: ...us situation which if not avoided may result in minor or moderate injury NOTICE Notice Indicates a hazardous situation which if not avoided may result in property damage WARNING Injury from heavy load...

Страница 14: ...te them as instructed on the screen WARNING Personal injury or damage to the analyzer due to contact with instrument mechanism Do not touch any parts of the instrument other than those specified Durin...

Страница 15: ...mmediately with water and apply a dis infectant Consult a physician WARNING Injury through reagents and other working solutions Direct contact with reagents cleaning solutions or other working solutio...

Страница 16: ...found to comply with the limits for Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reason able protection against harmful interferences when the equip...

Страница 17: ...ohazard material CAUTION Fatigue due to long hours of operation Looking at the monitor screen over an extended period of time may lead to fatigue of your eyes or body Take a rest for 10 to 15 minutes...

Страница 18: ...low connection to authorized external devices only o Ensure that all external devices are protected by appropriate security software o Ensure that access to all external devices is protected by approp...

Страница 19: ...se For details see document CedexBio Sys tem License Notes The portions of used open source software are part of MIKRAPs CPU X168LCD NET board support package and may be obtained from Mikrap AG info m...

Страница 20: ...re contaminated or not If contaminated treat in the same way as the instrument Legal liability It is the users sole responsibility to determine whether they activate the TRL Check feature and if they...

Страница 21: ...g out maintenance procedures When working with the instrument be sure to observe both the safety labels on the instrument and the safety notes in the Operator s Manual A This label on the electrode bl...

Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...

Страница 23: ...ex Bio instrument Overview A 17 Principles of operation A 19 User interface A 20 Wizards A 21 Daily operation A 22 Overview A 22 Reagent and diluent handling A 24 Calibration A 26 Calibration type A 2...

Страница 24: ...Roche Diagnostics A 16 Operator s Manual Version 3 1 2 Introduction to the instrument Cedex Bio System Table of contents...

Страница 25: ...tenance improves up time and reduces maintenance costs o Adaptable user interface The built in color touchscreen process driven software and reagent and sample barcode entry adapts to users of differe...

Страница 26: ...ap covers wash station ISE tower tubing B Main cover covers rotor reagents cuvettes photometer unit C Main switch D Transfer head holds probe E Rear service flap covers computer boards power supply de...

Страница 27: ...and tubing are flushed with water and cleaner 3 Wait The fluid needs to reach the prescribed temperature Such a phase can last several cycles During the wait cycles activities for other tests are per...

Страница 28: ...details on orders and Utilities for administration tasks e For details on the user interface see Chapter 4 Software The following is an example of a screen It contains the full range of display items...

Страница 29: ...most tasks that involve more than one step such as exchanging reagent or other fluid bottles you are supported by wizards Prepare wizard The Prepare wizard guides you through the tasks that need to be...

Страница 30: ...Navigation With wizard As individual steps 1 Starting the system 1 Switch on the system 2 Logging on the system Overview Logon 3 Preparing the system Start the Prepare wizard Overview Prepare 1 Check...

Страница 31: ...rforming controls Performing Default QC 1 Start the wizard Overview Order 2 Select a control and place the tube Repeat until there are no controls left on the screen n a 3 Start the QC measurement 4 V...

Страница 32: ...ce Result Review Choose Not Sent to Host 4 Start the End Shift wizard Overview End Shift 5 Perform the daily backup Utilities Export Database 6 Export the full results Utilities Export Results 7 Clean...

Страница 33: ...efine more than one reagent set for a test but only one can be active A reagent set is defined as soon as its first bottle is loaded From this moment on whenever you remove or replace a reagent you do...

Страница 34: ...valid for the calibrated set only They can be generated from any reagent set Lot calibration Lot calibration results are valid for the reagent set they were calibrated with and for all subsequent reag...

Страница 35: ...Each Set and Interval You perform calibration whenever a new reagent is loaded and when the interval has expired The interval starts again whenever you calibrate a reagent set because the interval ha...

Страница 36: ...ation result storage The current and up to five obsolete calibration results are stored on the system If there are more than five obsolete calibration results the oldest obsolete calibration results a...

Страница 37: ...is is typically done both for the normal and the pathological analyte concentration When QC is due With regards to when it needs to be performed QC is divided in the following types o QC After Cal The...

Страница 38: ...able flags list on page B 160 If you ignore a QC result you exclude the result from further QC result calculations such as QC History statistics Sample handling You can place up to eight sample tubes...

Страница 39: ...ou can remove sample tubes as soon as pipetting is complete Order handling Order Mode The order mode reflects the way in which you organize the tests on the test selection screen Choose Easy if the re...

Страница 40: ...guides you through the process of assigning the tests to the sample and placing the sample tube on the instrument STAT orders STAT short turn around time orders are handled in the same way as routine...

Страница 41: ...mple result with the exception that they cannot be accepted by the user They are automatically accepted if all their constituent results are accepted Repeating and rerunning tests If a result is flagg...

Страница 42: ...tion Performing all due maintenance actions during the daily Prepare or End Shift phase ensures that routine operation does not have to be interrupted for performing maintenance actions System status...

Страница 43: ...rming tests Diagnostics A system diagnostics action is being performed The system is not available for performing tests Operating Processing is in progress Power Up After switching on the system perfo...

Страница 44: ...Roche Diagnostics A 36 Operator s Manual Version 3 1 2 Introduction to the instrument Cedex Bio System System status...

Страница 45: ...w A 44 Sample area A 44 Fluid system A 45 Probe and syringe A 46 Syringe assembly A 46 Wash station A 47 External fluid connectors A 47 Transfer unit A 48 Fluid containers A 50 Waste containers A 50 W...

Страница 46: ...Roche Diagnostics A 38 Operator s Manual Version 3 1 3 Hardware Cedex Bio System Table of contents Absorbance photometer A 61 Connectors A 63 Fuses A 64 Technical specifications A 65...

Страница 47: ...and switch off the instrument before you open this cover Left service flap Provides access to o Wash station o Tubing o Internal waste tank o Initialization plate Handle as instructed during maintena...

Страница 48: ...omputer boards o Power supply o Transfer unit o Degasser This flap should be opened by service representatives only Shut down the system and switch off the instrument before you open this cover and re...

Страница 49: ...e main cover User intervention is required for example you are expected to place or remove a bottle The system is performing some action Do not handle the cover An acoustic signal is sounded when the...

Страница 50: ...rm messages There is at least one unconfirmed alarm message You need to deal with it as soon as possible There is at least one unconfirmed alarm message You need to deal with it immediately processing...

Страница 51: ...for loading removal pipetting and measuring Transfer unit Pipettes sample reagent and other fluids from their source to target containers such as cuvettes or the wash station Photometer unit Contains...

Страница 52: ...tubes This area is also used for placing calibrators controls and auxiliary fluids Display The touchscreen provides the user interface for controlling and managing the Cedex Bio instrument Front USB p...

Страница 53: ...tem water and waste The fluid system also delivers the correct amounts of fluids to the cuvettes A Wash station B Transfer head with probe C Pumps D Degasser E Syringe assembly F Internal waste tank G...

Страница 54: ...esh water and cleaner e See maintenance action Clean the probe manually on page C 12 See maintenance action Deproteinize the probe on page C 10 See Replacing the probe on page B 125 Syringe assembly T...

Страница 55: ...it in and outside Next water is pumped through the probe to flush away the cleaner The wash station is connected by tubing to the internal waste container External fluid connectors The three external...

Страница 56: ...eaning actions The following figure shows the major parts of the transfer unit Transfer head The transfer head moves horizontally along the X axis the probe moves up and down Z axis and it performs a...

Страница 57: ...ottom detection Because such a probe cannot pierce a bottle cap all bottles must be placed on the instrument with their caps removed Level detection A sensor detects when the probe enters a fluid On t...

Страница 58: ...t cooler condensation o ISE unit Fluid Container s Position Sample Tube Sample area Control Tube Sample area Calibrator Tube Sample area Diluent Reagent bottle Reagent disk System cleaner External cle...

Страница 59: ...connecting the external fluid containers on page B 127 See Checking the status of the external fluid containers on page B 15 See maintenance action Clean the water and waste containers on page C 16 Wa...

Страница 60: ...r of cleaning and pipetting actions that were performed When delivered the bottle has a white cap During installation this is replaced by a red cap with tubing attached to it tubing adapter The cleane...

Страница 61: ...consists of up to three reagents You always load and replace all bottles of a set Chimneys are bottle inserts that reduce evaporation For reagents that are especially sensitive to concentration chang...

Страница 62: ...tubes are placed centrally on the primary tubes and that they rest fully on them NOTICE Probe damage due to not removing primary tube caps The probe is not designed to pierce tube caps It can get dam...

Страница 63: ...e cuvette segments to the cuvette port where you can load or remove them You handle one segment at a time Cuvette segments are placed in the cuvette ring of the rotor e See Preparing cuvettes on page...

Страница 64: ...reagents on page B 22 Rotor The rotor provides the following features o Space for up to 27 reagent bottles on the reagent disk o Space for up to 60 cuvettes in the cuvette ring o A cooled environment...

Страница 65: ...gned to be handled as one unit including the bottles When not used on the instrument the reagent disk is placed in a container and stored in a refrigerated place e See Preparing the reagent disk on pa...

Страница 66: ...d tab Reagent disk container For storage outside the instrument the reagent disk is placed in a container This reduces evaporation of reagents and prevents their contamination Reagent cooler The reage...

Страница 67: ...ee maintenance action Clean reagent disk and sample area on page C 14 Barcode scanner A hand held barcode scanner is used for reading barcoded labels A Cuvette ring B Cuvette segment Figure A 28 Cuvet...

Страница 68: ...els Reagent bottle barcode On the reagent bottles barcode of the PDF417 format is used The barcode contains the following information o Part ID o Lot number o Expiration date o Reagent volume o Serial...

Страница 69: ...ling printer paper on page B 122 See To clear the paper jam on page D 46 Absorbance photometer The Cedex Bio main instrument uses the absorbance photometry measuring method Absorbance photometer The m...

Страница 70: ...rview Wavelengths for the absorbance photometer For each cuvette the absorbance photometer measures light intensity at 12 different wavelengths Measurement principles e See Principles of operation on...

Страница 71: ...ta and communication connectors C Fluid connectors D Front USB port not shown E Serial communication connector F Barcode scanner connector G Maintenance connectors Do not remove the cable H USB 2 conn...

Страница 72: ...he connector panel e See To change the mains fuses on page D 47 See To change a low voltage fuse on page D 49 WARNING Electrical shock by electronic equipment Do not attempt to work in any electronic...

Страница 73: ...Installation category II IEC 61010 1 Main fuse T6 3 A H 250 V Low voltage fuses T3 15 A L 250 V Battery Lithium 3 6 V 2 3 Ah SL 360 S Line voltage 100 240 V 10 Line frequency 50 Hz 5 and 60 Hz 5 Suppl...

Страница 74: ...Barcode Barcode 2 D format PDF417 Number of tests 50 200 tests depending on the test Segments of 10 cuvettes Manual insertion and removal of segments Single use Cuvettes are for single use only Stati...

Страница 75: ...A 71 Screen layout A 72 Display items A 73 Color concept A 74 Workflows and wizards A 74 Navigation A 74 Working with the user interface A 75 Adjusting the touchscreen A 75 Scrolling A 75 Expanding an...

Страница 76: ...y A 101 Calibrations list A 103 Lot data A 104 Lot list A 105 Loadlist A 106 Worklist A 106 Utilities tab A 107 Configuration A 108 Maintenance A 109 Applications related functions A 110 Applications...

Страница 77: ...kly find the required information and to locate the various display items Buttons Press a button to open and close a screen or to start a func tion Colors The color of an item on the screen points to...

Страница 78: ...id data A The status line displays the system status B Tabs represent the major work areas You can switch to any of them any time C The headline characterizes the content or function of the screen If...

Страница 79: ...tells you that a new screen will be displayed when you press the button a triangle in the bottom left corner that the current screen will be closed Global action button The global action buttons are...

Страница 80: ...he workflow is realized as a sequence of screens a so called wizard Cedex Bio wizards do not usually force you to perform a task at a certain stage they just make your work easier Navigation Moving fr...

Страница 81: ...ead to the expected result e See Adjusting the touchscreen on page B 131 Scrolling If not all text or all list elements fit on one screen or display area use the scrolling function to display the hidd...

Страница 82: ...levels of entries are marked with a To expand a list 1 Select a list item marked with 2 Press again or press 3 Use the scrollbar if required to display the items you are interested in a To collapse a...

Страница 83: ...Z Switch to the uppercase alphanumeric keyboard Alphanumeric keyboards Switch to the lowercase keyboard Switch to the uppercase keyboard Delete the last character displayed in the text line Switch to...

Страница 84: ...rd Delete the last character displayed in the text line Press to insert a space Using the filter function In many lists you can apply a filter that is you can select predefined criteria for generating...

Страница 85: ...layed again It now contains only the entries that comply with the criterion you just applied by pressing its button A In the status line the screen is indicated where you pressed A After you have appl...

Страница 86: ...cting the kind of data you want to print In these cases the print button is marked with a triangle in the top right corner a To print information 1 Press If filter criteria are available a screen is d...

Страница 87: ...a permanent feature of all screens a To display the Online Help 1 Press A The Help button is always in the top right corner of the screen Figure A 43 Permanent Help button A A The Legend tab describes...

Страница 88: ...uring operation is reported in the alarm monitor Message screen Message screens are displayed automatically as soon as the message is generated Read the message and press to close the screen Alarm mon...

Страница 89: ...Acoustic signal An acoustic signal is sounded when an alarm is generated You can adjust the volume Utilities Configuration System Volume a To display alarm messages 1 Press 2 Use to limit the number o...

Страница 90: ...e tabs and point out the main tasks you can perform on them Overview tab The Overview tab is your main work area when performing the daily routine tasks Show details about the order for this sample ST...

Страница 91: ...egment is available The last available cuvette segment is in use No cuvettes are available System Status The System Status button displays both the icon and the color of one of the buttons of the unde...

Страница 92: ...rs when you do so All tests are completed but not yet accepted All remaining tests are blocked for one of the following reasons o There is not enough sample fluid o The sample is not identified There...

Страница 93: ...displayed depends on how your Cedex Bio instrument is integrated in your laboratory infrastructure barcodes host connection Identifying samples f Overview Order Order Interval QC Order Default QC Typ...

Страница 94: ...ratios precede the tests and they adopt the color of their tests The test is on board and ready for use The test has already been pipetted The test is blocked The expiration date of the test has passe...

Страница 95: ...der to be processed irrespective of what routine orders already exist Existing STAT orders are finished first Tests Displaying the test overview f Overview Tests Tabs The tabs are displayed if you wor...

Страница 96: ...together with information on their status Handle the reagent disk that is currently on board Displaying test details f Overview Tests test button The color of the text indicates whether you need to r...

Страница 97: ...a shift End shift f Overview End Shift Start the End Shift wizard to perform the tasks necessary for ending the shift Cuvette status f Overview The six cuvette segments are represented by buttons The...

Страница 98: ...d diluent or cleaner is not on board Display detailed information on the status of this test Load a reagent set The number to the left of the button indicates the number of free positions on the disk...

Страница 99: ...to the reference electrode After installing a new electrode the Electrode Service maintenance action is due Cal Ref Display detailed information on the status of the ISE fluid bottle No fluid registe...

Страница 100: ...the status of hardware items and on IDs of installed software Status of the instrument The main cover is open Temperature status for reagent cooler and cuvette ring The temperature is outside the acce...

Страница 101: ...rocessed started tests may have to be restarted after refill Waste button Press to confirm that you have emptied the external waste container The waste container was last emptied less than one day ago...

Страница 102: ...iew and validate active QC results QC History View QC results on individual screens of the current or the previous calendar month or those generated before the previous month Calibrations View validat...

Страница 103: ...the button name o Detail of Order The currently selected order with its results Change the selected order Define a new order Delete orders You can choose what kind of orders should be deleted o List A...

Страница 104: ...s line contains sample information This line is selected S column Status This result has not yet been accepted Print results You can choose one of the following options o List All results in the list...

Страница 105: ...etransmit Display details of the selected result Delete results You can choose what kind of result should be deleted o List All results in the list If a filter was applied the list would contain only...

Страница 106: ...esult if no flag was generated Flag with highest priority if flags were generated Order status if the control measurement has not been performed yet Previous Lot indicates that controls of more than o...

Страница 107: ...are grouped first by test name then by control Last Measurement column Date and time of the most recent result Previous Lot indicates that control was performed after a QC lot change Print QC results...

Страница 108: ...elevant events QC lot change reagent set change QC target value change precede the results Month button Display the results of the month indicated on the button Old Display the results generated befor...

Страница 109: ...played In all other cases the order status is displayed Print calibration results You can choose one of the following options o List If a filter was applied the list would contain only the results tha...

Страница 110: ...bration result Apply filter criteria to the calibration results list You can choose one of the following criteria o All o Current Lot data f Workplace Lot Data Choose the kind of material for which yo...

Страница 111: ...collapse the list With a value selected Change the value With a calibrator selected Add a lot With a test selected Assign a test to the lot With a value selected Not active With a calibrator selected...

Страница 112: ...the loadlist Worklist The worklist shows all tubes currently placed on the sample area f Workplace Worklist Print the worklist A Number of due pipettings B Number of available pipettings Figure A 69...

Страница 113: ...xtra wash cycles o Display export and import the mapping table for test IDs of the Cedex Bio instrument and the laboratory information system Diagnostics Perform diagnostics actions Inventory Display...

Страница 114: ...again to expand the list and display the items The list is expanded Select the item marked with and press again to hide the items Print configuration data You can choose one of the following options...

Страница 115: ...t major maintenance day The defined maintenance interval has expired Perform this maintenance action now Perform the selected maintenance action A Maintenance action name B Due dates in ascending orde...

Страница 116: ...s of the installed applications and import and install new applications Extra Wash Cycles Handle and install extra wash cycles Host Codes Display export and import the mapping table for test IDs of th...

Страница 117: ...the list and display the items The list is expanded Select the item marked with and press again to hide the items Print the application list With an item selected that is preceded by or expand or col...

Страница 118: ...by or expand or collapse the list With any other item selected change the selected item Import a new extra wash cycle Delete the selected extra wash cycle Host codes f Utilities Applications Host Cod...

Страница 119: ...cted mixing rule Process sequence f Utilities Applications Process Sequence Print the process sequence list In the Priority list move the item up In the Priority list move the item down Move the selec...

Страница 120: ...stic action selected Perform the selected action Diagnostics actions are designed for use by Roche representatives or when instructed by them to do so Therefore these functions are not described in th...

Страница 121: ...han 30 days ago are automatically deleted from the Inventory list provided there is still a valid set for the same fluid on board If this were not the case the set that was loaded last would remain in...

Страница 122: ...s Import Application Import application data Software Import software updates Database Import the complete content of the database Certificate Import certificates that authenticate reagent barcodes EW...

Страница 123: ...ress again to hide the items Print the user data of the selected user With an item selected change the selected item With any other item selected expand or collapse the list Define a new user Delete t...

Страница 124: ...all systems and modules that are currently not ready Sampling stop Finish the current pipetting action but do not start a new one You can restart processing by pressing Stop printing Stop the current...

Страница 125: ...intervention is required for example you are expected to place or remove a bottle The system is performing some action Do not handle the cover An acoustic signal is sounded when the cover is open whil...

Страница 126: ...confirmed alarm messages There is at least one unconfirmed alarm message You need to deal with it as soon as possible There is at least one unconfirmed alarm message You need to deal with it immediate...

Страница 127: ...e Delete the selected item Cancel Abort the operation Do not use this button while the system is performing control or calibration tests this might lead to data loss Close Close the screen Print Print...

Страница 128: ...k Insert Bottle Place a reagent bottle Remove Bottle Remove a reagent bottle Mixing Perform reagent mixing Replace Cuvette Confirm replacement of the cuvette segment Remove Cuvette Confirm removal of...

Страница 129: ...he first item in a list To Bottom Select the last item in a list Forward Open the next screen in a wizard Next Step Back Open the previous screen in a wizard Previous Step Table A 14 Navigation button...

Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...

Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...

Страница 132: ......

Страница 133: ...em B 11 Preparing the system B 12 Short guide B 13 Starting the Prepare wizard B 15 Checking the status of the external fluid containers B 15 Performing maintenance actions B 19 Preparing the reagent...

Страница 134: ...t QC B 70 Performing interval QC measurements B 71 Validating QC results B 72 Interpreting the QC history B 74 Performing reagent mixing B 76 Finishing the shift B 77 Safety information B 77 Short gui...

Страница 135: ...Contact with samples containing material of human origin may result in infection All mate rials and mechanical components associated with samples of human origin are potentially biohazardous o Be sure...

Страница 136: ...y Roche Diagnostics o Start all maintenance actions on the screen Do not perform maintenance actions with out the assistance of the user interface o Carefully follow the procedures specified in the Op...

Страница 137: ...he Prepare wizard Overview Prepare 1 Check the external fluid containers Overview 2 Perform the maintenance actions that are due Utilities Maintenance 3 Load the reagent disk Overview 4 Check the reag...

Страница 138: ...ing controls Performing Default QC 1 Start the wizard Overview Order 2 Select a control and place the tube Repeat until there are no controls left on the screen n a 3 Start the QC measurement 4 Valida...

Страница 139: ...Choose Not Sent to Host 4 Start the End Shift wizard Overview End Shift 5 Perform the daily backup Utilities Export Database 6 Export the full results Utilities Export Results 7 Clean up the database...

Страница 140: ...rument Preconditions Before you switch on the instrument make sure the following preconditions are met o All covers are closed o The external fluid containers are connected o The sample area is empty...

Страница 141: ...pper and lower case As soon as you type the first characters of your user name the system looks for a name that starts with these letters and if it finds one displays the complete name 3 Press A scree...

Страница 142: ...daily preparation tasks is to follow the Prepare wizard Preventing intervention during sample processing Performing conscientiously all preparation steps reduces considerably the chances of you having...

Страница 143: ...he waste container if required and confirm by pressing the waste button 3 Replace the cleaner bottle if required and con firm by pressing the cleaner button 4 Press to proceed to the next stage in the...

Страница 144: ...les of the set 11 Press to proceed to the next stage in the Prepare wizard 6 Prepare the cuvettes Replace all red cuvette segments 1 Press the cuvette button 2 Open the main cover 3 Replace the cuvett...

Страница 145: ...ayed on the same screen Make sure that you have read and understood section Safety information on page B 5 The following warning messages in particular are relevant o Injury through reagents and other...

Страница 146: ...he waste container on page B 17 There is no water left A fluid sensor detected that there is no water coming from the water container You need to refill the water now No new tests can be processed sta...

Страница 147: ...place it on a firm and even surface o Place the spare container on the rack o Check that the tubing adapter is properly inserted o Empty the removed waste container Treat the fluid as biohazardous wa...

Страница 148: ...rm that you have exchanged the cleaner bottle Level monitoring for the cleaner bottle is based on the number of cleaning and pipetting actions that were performed The counter is reset when you press t...

Страница 149: ...e by when they should be performed Use these dates for planning the maintenance actions for example for ordering the required materials Interpreting the colors Make sure that you have read and underst...

Страница 150: ...on 5 Follow the instructions on the screen 6 Perform the next maintenance action that is due 7 When you have finished performing maintenance actions press to proceed to the next stage in the wizard WA...

Страница 151: ...oler The system automatically detects that a disk was inserted A screen is displayed asking you to close the main cover 4 Close the main cover At this stage the system identifies the disk A screen is...

Страница 152: ...declining reagent quality Due to the declining quality of reagents the quality of measurement may decline This may cause incorrect results Make sure to use reagents whose expiration date has not passe...

Страница 153: ...tle sets that are defined on any of the disks used on this Cedex Bio instrument including sets that were removed from the disk as long as they were not empty or their expiration date has not passed ye...

Страница 154: ...ration There are fewer than 10 of tests left for this set Load a new reagent set as soon as pos sible The expiration date has passed A QC is due or its result has not been accepted Perform QC as soon...

Страница 155: ...placing a reagent set try to use one of the same lot as the old one This way if you work with the calibration sequence Each Lot and Interval you avoid having to perform a calibration Removing a reagen...

Страница 156: ...nother bottle belonging to the set The system moves this bottle to the reagent port A screen is displayed asking you to remove the bottle 1 Remove the bottle 2 Press to confirm the removal 3 Start add...

Страница 157: ...t inserted a reagent set you may have to perform the initial calibration and QC Time limit for placing reagent bottles o The system assumes that you place the bottle that you just scanned o You need t...

Страница 158: ...s Checking the cuvette status 1 Check the color of the cuvette segment buttons You can refresh the cuvette screen by pressing If there were pending orders this would also start their processing Interp...

Страница 159: ...hat require mixing Within the Prepare phase a screen is displayed that lists all reagent sets for which a mixing interval is defined a To perform the mixing 1 Select the reagent sets you want to mix I...

Страница 160: ...status Make sure that you have read and understood section Safety information on page B 5 The following warning messages in particular are relevant o Injury through reagents and other working solution...

Страница 161: ...his test A Press to cancel the selection of all tests that need calibrating B Press to cancel the selection of all tests that need calibrating within the forecast period Figure B 8 B A Color Meaning P...

Страница 162: ...results 8 Validate the results e See Validating calibration results on page B 64 A Position on the sample area where to place the calibrator B Lot number C Calibrator with highest concentration D The...

Страница 163: ...operation Preparing the system End of Prepare phase The instrument is now ready for analyzing samples You may want to perform QC before analyzing samples e See Performing QC on page B 68 When performi...

Страница 164: ...ubbles or foam may cause level detection errors and air pipetting Consequently incorrect results may be generated Take adequate care when preparing the samples Incorrect results due to insufficient fl...

Страница 165: ...Select the tests ratios and profiles This step is not required if the workflow setting Working Mode is Host Utilities Con figuration Workflow 1 Press a tab to select the test panel if required 2 Press...

Страница 166: ...orkplace Results 2 Validate the result You can accept the result or perform the test again 6 Remove the samples 1 Press the Overview tab 2 Check the status of the sample tube button that corresponds t...

Страница 167: ...ups of tests You can define up to 20 tests and profiles on each panel tab e See Assigning tests to test tabs on page B 119 Sample ID Handling The sample ID is an identifier of up to 23 alphanumeric ch...

Страница 168: ...the sample area WARNING Incorrect results due to insufficient fluid Insufficient fluid may lead to inaccurate pipetting and consequently to incorrect results Always fill the tubes with enough fluid t...

Страница 169: ...kflow setting Working Mode is Host this screen will not be displayed Utilities Configuration Workflow If Do this You work with sample barcodes Use the barcode scanner to scan the bar code You do not w...

Страница 170: ...screen for selecting the tests would be displayed instead 7 Do one of the following The test is blocked for one of the following reasons o The calibration failed o Initial calibration is required o Fo...

Страница 171: ...rders e See Repeating tests on page B 55 See Rerunning tests on page B 56 Defining STAT orders Short turn around time STAT orders are defined in the same way as routine orders a To define a STAT order...

Страница 172: ...ully processed You can add further tests to an order any time The process of changing an order depends on whether the sample tube is still on the instrument or not Making a change to an order when the...

Страница 173: ...is no longer on the instrument A message is displayed asking you to place the sample o Place the sample on the sample area The orders list is displayed again 4 The system reschedules the order and pr...

Страница 174: ...n progress the orders are processed according to sequence in which they were defined Repeats and reruns of routine orders are performed before routine orders Monitoring the analysis progress During op...

Страница 175: ...e sample is not identified 1 Remove the sample 2 Scan the sample barcode or type the same sample ID as before 3 Reinsert the sample on any posi tion Removing a sample and then placing it again means d...

Страница 176: ...ple tube 3 Take appropriate action Checking the instrument status buttons a To check the instrument status 1 Press the Overview tab A Test name B Press to validate the results C Press to add tests to...

Страница 177: ...Replace the cuvette segments System Status The system status button displays both the icon and the color of one of the buttons of the underlying system status screen The icons are first prioritized by...

Страница 178: ...atus has changed to Standby LED Meaning Possible action No color off There are no unconfirmed alarm messages No action is required Yellow There is at least one unconfirmed alarm message User intervent...

Страница 179: ...esult are considered not measured Recover When in Standby or Stopped status Initialize all systems and mod ules that are currently not ready Sampling stop Finish the current pipetting action but do no...

Страница 180: ...sample when its tube button is green e See Checking the sample tube status on page B 44 All tests are accepted All tests are pipetted 1 1 If you remove a sample tube before the pipetting is complete t...

Страница 181: ...f flagged results depends on the configuration Utilities Configuration Result Handling The following table shows how e See Editing the acceptable flags list on page B 160 There are two main approaches...

Страница 182: ...s activated the result might be displayed for example as 0 5 and the Test Rng flag is added This means that the measured value is below the defined TRL value 2 Select the result if there is more than...

Страница 183: ...accept results that are not flagged In addition it can be set up to accept results with flags that are marked in an editable list of flags that should be ignored which is particularly suitable if the...

Страница 184: ...ng the same dilution e See Repeating tests on page B 55 o Press Rerun to re perform the test using a predefined different dilution e See Rerunning tests on page B 56 o Press Accept to accept the resul...

Страница 185: ...n page B 52 e If the sample is no longer on board see To validate sample results from the results list on page B 53 2 Press Repeat The system automatically creates a new order and selects the test It...

Страница 186: ...results from the results list on page B 53 2 Press Rerun The system automatically creates a new order and selects the test The dilution factor is part of the test definition and therefore automatical...

Страница 187: ...editable list of flags that should be ignored Utilities Configuration Result Handling e See Result handling on page B 165 See Editing the acceptable flags list on page B 160 a To accept a result 1 Va...

Страница 188: ...ll the filter criterion and the filter criterion would be part of the button name Press Detail of Order to print all results of the associated order For repeated or rerun orders all results of this or...

Страница 189: ...nflammation caused by reagents Direct contact with reagents may cause skin irritation inflammation or burns When handling reagents be sure to wear protective equipment and observe the cautions given i...

Страница 190: ...brations 2 Press to select all tests that will require cal ibrating within the forecast period 3 Check and if required print the placement list 4 Prepare and place the calibrators according to the pla...

Страница 191: ...creen is displayed that lists all tests and their currently valid calibration 2 Press A screen is displayed that shows all active tests Interpreting the colors A Press to select all tests that now nee...

Страница 192: ...sts are selected whose calibration will be due within the forecast period The currently active lots and sets are used for the calibration 2 Press You want to calibrate the current set 1 Select the tes...

Страница 193: ...rating a new reagent set 6 Choose Workplace Calibrations 7 Press 8 Select the test 9 Press If a screen is displayed for selecting the calibration type choose Lot Master 10 Prepare the calibrators 11 P...

Страница 194: ...n The effect of flagged calibration results depends on the configuration e See Result handling on page B 165 See Editing the acceptable flags list on page B 160 Configuration setting Effect Cal Auto A...

Страница 195: ...splay all or only the current results Figure B 22 A E F G H B C D Item Value Interpretation U Use CU Current The calibration results of this set are currently used SB Standby The calibration results o...

Страница 196: ...ed an accepted result When you press Use Old a copy of the old result is made and a new entry is displayed in the calibration results list Note that the intervals are reset as if a new calibration res...

Страница 197: ...cepting calibration results Calibration results will be used by the system only when you have accepted them There can be only one unaccepted calibration result for a test If there already exists a not...

Страница 198: ...y QC orders are automatically defined as soon as you identify a control An order is defined for all tests for which this control is defined as the Default QC provided the test is currently active on t...

Страница 199: ...rements that are due 1 Choose Overview Order or Workplace QC Status 2 Press 3 Press 4 Select a control and place the tube Repeat until there are no controls left on the screen 5 Press Perform a single...

Страница 200: ...control button A screen is displayed asking you to place the selected control 5 Place the control tube on any free position on the sample area The system registers the position and automatically defin...

Страница 201: ...test button 3 Press A screen is displayed that contains a button for each of the controls that are required 4 Prepare the controls A Press to select all tests with due QC Press again to cancel the sel...

Страница 202: ...erview tab is displayed 8 Press to start processing the control orders Validating QC results Flagged QC results If flags are generated you must determine their cause and decide whether to accept the r...

Страница 203: ...s not been accepted yet I The result was ignored The result has not been transmitted yet B Test name C Control ID D Result if no flag was generated Flag with highest priority if a flag was generated O...

Страница 204: ...Interpreting the QC history The QC history provides on individual screens information on the QC results of the current and the previous calendar month as well as on results that were generated before...

Страница 205: ...to print the graphic L Press to display the results in a table M Press to display the results of the previous month N Press to display the results of the current month O Press to display the results t...

Страница 206: ...ication definitions and cannot be changed by the user The system checks every 30 minutes for reagent sets that require mixing e For mixing during the Prepare phase see Performing reagent mixing on pag...

Страница 207: ...ment Safety information Make sure that you have read and understood section Safety information on page B 5 The following safety messages in particular are relevant Warning messages o Injury through re...

Страница 208: ...d Shift wizard 1 On the Overview tab press the End Shift but ton 3 Perform data backup 1 Press 2 Insert the USB stick 3 Press 4 Select the directory 5 Press 6 Press to proceed to the next stage in the...

Страница 209: ...ette segments 4 Close the main cover 5 Press to proceed to the next stage in the End Shift wizard 8 Remove the reagent disk 1 Choose Overview 2 Press 1 Open the main cover 2 Lift the disk from the ins...

Страница 210: ...screen is displayed that contains all sample results of the day 2 Choose Not Accepted 3 Press A screen is displayed that contains all sample results that have not been accepted yet 4 Select a result...

Страница 211: ...alidating results is displayed 6 Press Retransmit The list with the results that were not transmitted successfully is displayed again The result you just retransmitted should no longer be present 7 Se...

Страница 212: ...For information on restoring the database see Importing a database on page B 113 a To perform the daily backup 1 Press A screen is displayed asking you to insert the USB stick 2 Insert the USB stick A...

Страница 213: ...v files a To export the results 1 Press A screen is displayed asking you to insert the USB stick 2 Insert the USB stick A screen is displayed showing the directory structure of the USB stick 3 Select...

Страница 214: ...obsolete if the empty set was removed or if the set was removed more than 30 days ago 2 Press to confirm the deletion The results are deleted 3 Press to proceed to the next stage in the End Shift wiz...

Страница 215: ...maintenance actions 1 Follow the instructions given in To perform maintenance actions on page B 19 2 Press to proceed to the next stage in the End Shift wizard A screen is displayed that shows the st...

Страница 216: ...er and place it on a clean surface 2 Follow the instructions given in To check the water container on page B 16 Make sure that you have read and understood section Safety information on page B 5 The f...

Страница 217: ...gent disk container in a refrigerated place Make sure that you have read and understood section Safety information on page B 5 The following warning messages in particular are relevant o Injury throug...

Страница 218: ...ng off the instrument Preconditions The system must be in Standby status a To shut down the system and switch off the main instrument 1 Press 2 The Stop screen is displayed 3 Press Shutdown to shut do...

Страница 219: ...ear a beep and release the trigger If the reading was successful a screen is displayed asking you to place the bottle on the reagent disk WARNING Loss of sight The intense light of the LEDs may severe...

Страница 220: ...barcode 5 Wait until you hear a beep and release the trigger If the reading was successful a screen is displayed asking you to place the item Reading sample tube barcodes a To use the barcode scanner...

Страница 221: ...lot definitions B 98 Calibrator and QC lots B 98 Reagent lots B 99 Viewing the lot details of fluids that were used B 100 Adding a new lot B 101 Exporting data B 104 Exporting the database B 104 Expor...

Страница 222: ...s Manual Version 3 1 6 Special operations Cedex Bio System Table of contents Replacing the probe B 125 Connecting and disconnecting the external fluid containers B 127 Adjusting the touchscreen B 131...

Страница 223: ...To delete orders 1 Choose Workplace Orders A screen is displayed that lists all currently defined sample orders 2 Do one of the following The orders are deleted You cannot retrieve them any more from...

Страница 224: ...should be accepted and printed or sent to the host before deleting a To delete sample results 1 Choose Workplace Result Review A screen is displayed that lists all results 2 Do one of the following Th...

Страница 225: ...results become obsolete when the empty set is removed or if it was removed more than 30 days ago This section describes how to proceed if you need to delete results outside the normal daily routine e...

Страница 226: ...d if you need to delete QC results outside the normal daily routine e For deleting QC results at the end of a shift see Cleaning up the database on page B 84 Definition item Configuration path Effect...

Страница 227: ...rom the system If Do this You want to delete all QC results 1 Choose All 2 Press to confirm the deletion You want to delete all accepted QC results 1 Choose All Accepted 2 Press to confirm the deletio...

Страница 228: ...rent lot definitions Calibrator and QC lots a To view calibrator or lot data 1 Choose Workplace Lot Data 2 Press either Controls or Calibrators A screen is displayed that lists all installed lots 3 Se...

Страница 229: ...s You can view the lot number in the reagent set details a To check the status of an individual reagent set 1 Choose Overview 2 Press a reagent set icon A screen is displayed that contains information...

Страница 230: ...ion 4 Press to close the dialog boxes until the Utilities tab is displayed Viewing the lot details of fluids that were used For sample calibration and QC results you can check to which lot the used fl...

Страница 231: ...rt With the exception of reagent and diluent lots they can be defined by scanning their barcodes only Preparation tasks Make sure that the controls and calibrators are assigned to the application e Se...

Страница 232: ...press 5 Type the expiration date Use the date format as indicated on the screen 6 Press to confirm the definitions If a lot with the same ID already exists on the system a screen is displayed asking...

Страница 233: ...ups you need to place on the instrument Use the information given on the package insert Exclude calibrators that use system water from this number See step 14 7 Press to confirm the definitions 8 Pres...

Страница 234: ...During a database export the full content of the database is copied to the USB stick The database data can be restored to the instrument if required e See Importing a database on page B 113 a To expo...

Страница 235: ...22 o Line terminator Standard Win Style CRLF ASCII Code 13 10 0x0D 0x0A These files can be processed with any spreadsheet program that can import csv files a To export the support data 1 Choose Utilit...

Страница 236: ...x 166 ID of the data set used App Application code Date Date when the result was calculated Time Time when the result was calculated SW Version Software version installed on the instrument Serial Seri...

Страница 237: ...rd number User User name SYS if Auto Accept is on Lot Lot number LS Value Calibration type Target value of corresponding calibrator L Lot calibration S Set calibration Unit Defined display unit Flags...

Страница 238: ...e calculated in Abs First Calculation Point Last Calculation Point Raw1 Raw40 Absorbance value of cycle 1 through 40 in Abs one cycle 18 s Test Short name of the application Ignored I if the QC result...

Страница 239: ...agent usage changed 192 Calibrator assigned valued changed 193 QC assigned valued changed 199 Other process event App Application code Date Date when the event took place Time Time when the event took...

Страница 240: ...rm messages and system logs a To export the log files 1 Choose Utilities Export Log Files A screen is displayed asking you to insert the USB stick 2 Insert the USB stick 3 Press to confirm the inserti...

Страница 241: ...vates the application on the instrument and so makes it available for use a To import application data 1 Do one of the following o Choose Utilities Applications Laboratory Parameters then continue wit...

Страница 242: ...d from the existing application The application definitions are replaced on the system by the new ones The application code exists and the short name does not exist for any application on the system T...

Страница 243: ...hen the import is complete the Overview tab is displayed and the system is in Standby status Importing a database Compatibility o If the version of the imported database is older than the one on the t...

Страница 244: ...to display all files and directories 7 Press to confirm the selection When the data are imported a message is displayed asking you to restart the system 8 Choose Restart to restart the system Bottle s...

Страница 245: ...cycle information is contained in a separate barcode on a barcode transfer sheet a To import mixing and extra wash cycle information 1 Choose Utilities Import EWC Mixing A screen is displayed asking y...

Страница 246: ...protective equipment Take extra care when working with protective gloves these can easily be pierced or cut which can lead to infection o If any biohazardous material is spilled wipe it up immediatel...

Страница 247: ...stick the label on the disk A 3 Using pliers break off the ID tab C D with the same number as the one printed on the label Loading the disk 4 On the system choose Overview A disk overview screen is d...

Страница 248: ...n is deemed empty When the first reagent bottle of a set is loaded the reagent set is defined From this moment on the reagents are handled as part of the set You no longer handle them as individual re...

Страница 249: ...y panel if there is space and to the first Full mode tab If there is not enough space on the first Full mode tab the test is added to the next tab that has space available a To name a test tab 1 Choos...

Страница 250: ...ent disks used on the Cedex Bio instrument You can delete bottle sets from the list this would be appropriate if you removed a set from the disk and did not intend to use it again a To delete a bottle...

Страница 251: ...Cedex Bio instrument including sets that were removed from the disk as long as they were not empty or their expiration date has not passed yet In the D column removed bottle sets are marked with R A...

Страница 252: ...red if the printer is out of paper 2 Scroll to display the Printer entry and read the text a To replace the printer paper 1 Make sure the printer is not currently printing 2 Open the printer panel 3...

Страница 253: ...t the paper in the slot in the printer panel and pull some through 8 Close the printer panel The system feeds some paper If the system ran out of paper during printing it will resume printing 9 On the...

Страница 254: ...n water from the reagent cooler a To remove water from the reagent cooler 1 Make sure the system is in Standby status 2 Choose Overview 3 Open the main cover 4 Remove the reagent disk Be sure not to t...

Страница 255: ...ing solutions may cause personal injury When handling such solutions exercise the precautions required for handling them observe the cautions given in the package insert and observe the information gi...

Страница 256: ...be assembly Treat it as biohazardous waste a To install the new probe 1 Carefully insert the probe in the probe holder and fasten the screw that fixes the tubing to the holder C 2 Reinstall the probe...

Страница 257: ...proteinize Probe maintenance action e See Deproteinize the probe on page C 10 Connecting and disconnecting the external fluid containers Before switching on the system make sure that all external flui...

Страница 258: ...onnect the waste tubing by pushing the plug firmly in the socket 3 Insert the tubing adapter in the container and push it down firmly a To connect the cleaner bottle 1 Place the bottle on the tray 2 T...

Страница 259: ...d containers 3 Screw the cleaner tubing to the connector on the instrument C Do not fasten the screw too tight 4 Push the protective sleeve firmly on its base on the instrument D 5 Insert the tubing a...

Страница 260: ...the connector away from the socket You want to disconnect the cleaner tubing 1 Pull out the protective sleeve B and move it along the tube far enough that you can grip the connecting screw 2 Turn the...

Страница 261: ...n for adjusting the screen is displayed 3 Press exactly in the center of the black cross When you have touched it properly it turns yellow and the next cross turns black 4 Do the same for all crosses...

Страница 262: ...rsion 3 1 6 Special operations Cedex Bio System Cleaning the touchscreen a To clean the touchscreen 1 Pour or spray some alcoholic solution on a tissue 2 Wipe the screen Exert as little pressure as po...

Страница 263: ...leting applications B 144 Defining the processing order B 145 Defining ratio applications B 146 Profiles B 148 Defining profiles B 148 Adding tests to a profile B 148 Removing tests from a profile B 1...

Страница 264: ...st codes B 160 Short guide to configuration definitions B 161 Workflow definitions B 161 Host definitions B 163 Result handling B 165 Calibration B 165 QC B 165 System B 166 Date and time B 167 Mainte...

Страница 265: ...t values are displayed Changing values If the change of a value results in one or several other values needing to be changed a wizard is started when you change the first value You recognize a wizard...

Страница 266: ...pressing its corresponding button 5 Type the new value or press a value button 6 Do one of the following If Do this You are in a wizard and want to change a further value 1 Press A screen is displayed...

Страница 267: ...ociates the application with lot data and auxiliary fluids such as calibrators and controls e To prepare an application on page B 140 a To import application data 1 Do one of the following o Choose Ut...

Страница 268: ...rom the existing application The application definitions are replaced on the system by the new ones The application code exists and the short name does not exist for any application on the system The...

Страница 269: ...application to temporarily make unavailable the test on the test panel All associated fluids and data remain unchanged You can activate the application later again and continue using the test a To ac...

Страница 270: ...name up to 5 alphanumeric characters if required The short name affects the display of the test name in the user interface for example test buttons or test names in the orders and results lists and i...

Страница 271: ...to confirm the definitions Control definitions 14 Expand the Control entry select Sequence and press 15 Define the sequence Choose No Interval for cases when for example you intend to perform QC as pa...

Страница 272: ...onfiguration System enter the TRL value lower limit of measuring range as defined in the relevant application documentation See also System on page B 166 30 Press to confirm the definitions Result Con...

Страница 273: ...ee Deleting QC results on page B 96 a To uninstall an application 1 Choose Utilities Applications Laboratory Parameters 2 Select the application 3 Press A confirmation screen is displayed 4 Press to c...

Страница 274: ...Choose Utilities Applications Laboratory Parameters 2 Select the application 3 Press The following table lists possible situations and how they affect the process of deleting A confirmation screen is...

Страница 275: ...test you want to be processed with high priority 3 Press to move it to the Priority list The tests are processed in the order displayed in this list When the tests contained in this list are processed...

Страница 276: ...tion 5 Type the short name and press The short name must be unique on the Cedex Bio system and consists of up to 5 alphanumeric characters It is used on the screens for example on the test selection s...

Страница 277: ...gh range will be used or not then press 18 Type the low limit then press 19 Type the high limit then press 20 Type the standard unit then press 21 Type the laboratory unit then press 22 Type the conve...

Страница 278: ...5 Type the short name up to 5 characters The short name will appear on the test selection screens 6 Type the long name up to 30 characters 7 Select the tests that should be included in the profile 8...

Страница 279: ...tion screen is displayed 7 Clear the selection of the tests you want to remove 8 Press to confirm the selection Deleting profiles e See Deleting applications on page B 144 Defining control definitions...

Страница 280: ...lly 1 Press 2 Type the name and press 3 Type the material code and press You find this code in Utilities Applications select a test Control 4 Type the Lot ID and press 5 Type the expiration date Use t...

Страница 281: ...system water from this number See step 14 7 Press to confirm the definitions 8 Press Add new test 9 Press A screen is displayed that shows all tests that have this calibrator defined for them and that...

Страница 282: ...lication definitions the user might want to change Item Values Comment Short Name Up to 5 alphanumeric characters The short name is used on the screens for exam ple on the test selection screen or the...

Страница 283: ...Set and Interval Due whenever a new reagent is loaded and when the interval has expired The interval starts again whenever you place a new reagent set You can turn off the interval check by defining i...

Страница 284: ...tically informs you about due QCs Interval Number of days hours for ISE applications Use 0 zero if the Sequence is set to No Interval For each of the controls Mat Code 0 999 Unique identification of t...

Страница 285: ...lts that are outside the limit are flagged High Limit Results that are outside the limit are flagged 1 3 Flag QC Fail On Off Choose On to ensure that all sample results are flagged that were generated...

Страница 286: ...he defined factor and offset are applied the final ratio result Table B 17 Laboratory correlation definitions for ratio applications Item Values Comment Valuation No Valuation No valuation is performe...

Страница 287: ...8 Press 9 Type the new password a second time 10 Press NOTICE Damage to data and programs due to unauthorized access to the Cedex Bio instrument User access to the Cedex Bio instrument is controlled b...

Страница 288: ...een actions that are due on Monday or an earlier day are marked in red actions that will fall due between Tuesday and Sunday will be marked in yellow The idea is to perform on the main maintenance day...

Страница 289: ...ho should perform routine tests o Press Supervisor for users who should perform routine tests as well as calibration and QC o Press Lab Manager for users who should perform all test actions and who sh...

Страница 290: ...6 Press Mapping the host codes Applications are delivered with a unique ID called code Laboratories often use their own codes for applications By mapping the codes used by the laboratory host codes t...

Страница 291: ...ose Utilities Host Codes 2 Press The host codes are set to values that are identical to the application codes By pressing again you can revert to the original values Short guide to configuration defin...

Страница 292: ...ify the sample Independent IDs You define the sample ID and order ID indepen dently of each other Grouped Sample ID Order and sample ID are identical For a given sample the same sample ID is used all...

Страница 293: ...are retrieved from memory or received at the other end of a network a new checksum is computed and matched against the existing checksum A non match indicates an error Order Query Mode Query mode all...

Страница 294: ...g normal daily operation The trace file can be exported as a log file Utilities Export Log Files file format htryyyymmd dhhmmss log System Device ID Up to 10 alphanumeric characters The ID is used on...

Страница 295: ...is on Cal Auto Accept On Off Results that do not contain a flag are automatically accepted Results with flags that are marked in a list of acceptable flags are accepted as well Cal Acceptable Flags Li...

Страница 296: ...t definition Sound Effects On Off Acoustic signals when handling instrument items e g clicking sound when typing Does not affect the acoustic signals for alarms and warnings Volume 0 100 Volume of any...

Страница 297: ...is used for time records for example with results Example 08 32 This format is for entry pur poses only It does not influence the format of dis played times Table B 26 Date and time definitions Item V...

Страница 298: ...Do this The definitions are available on a barcode 1 Read the barcode The definitions will be installed on the system You want to define the values manually 1 Press 2 Choose when in relation to the p...

Страница 299: ...t lists the currently defined extra wash cycles 2 Select and expand the extra wash cycle entry 3 Press 4 Choose On to activate the extra wash cycle Choose Off to deactivate the extra wash cycle 5 Pres...

Страница 300: ...extra wash cycle should be performed Trigger Bottle Code 0 999 Bottle set code of the fluid that triggers performance of the extra wash cycle If you use the pipetting type Sample it defines the bottl...

Страница 301: ...you want to print 3 Press A printing screen is displayed 4 Select whether you want to print the selected rule set or all defined rule sets The rules are printed on the built in printer You can import...

Страница 302: ...or Administrator user rights to perform this task a To delete a mixing rule 1 Choose Utilities Applications Reagent Mixing A screen is displayed that lists the currently defined mixing rules 2 Select...

Страница 303: ...8 General maintenance C 3 Maintenance C...

Страница 304: ......

Страница 305: ...s C 6 Maintenance actions and their intervals C 7 Maintenance actions C 8 Safety information C 8 Deproteinize the probe C 10 Prime the fluid system C 11 Clean the probe manually C 12 Clean reagent dis...

Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...

Страница 307: ...s can be performed any time Interval For most maintenance actions a fixed maintenance interval is defined You cannot change this interval This is the basis upon which the system calculates the date wh...

Страница 308: ...on 2 Select the maintenance action you want to perform The selected line turns blue 3 Press The first screen of the wizard for performing the selected maintenance action is displayed Usually this is t...

Страница 309: ...ction that was due its status will remain due and you need to fully re perform the action later If at all possible complete a maintenance action without interrupting it Every effort has been made to e...

Страница 310: ...e extra care when working with protective gloves these can easily be pierced or cut which can lead to infection o If any biohazardous material is spilled wipe it up immediately and apply disinfectant...

Страница 311: ...later o If at all possible complete a maintenance action without interrupting it o If you need to interrupt the maintenance action wait until you see confirmation on screen that the current step has...

Страница 312: ...with ISE Deproteinizer and one with Activator 2 Select the maintenance action Deproteinize Probe 3 Press The maintenance definition screen is displayed 4 Press A placement list is displayed 5 Place t...

Страница 313: ...nt in the fluid system your test results will be inaccurate Operator time Approximately 2 minutes Prerequisites The system must be in Standby status Tools and materials required m Purified water a To...

Страница 314: ...maintenance definition screen is displayed 3 Press The system deactivates the transfer motors 4 Remove the transfer head cover Press the release buttons on both sides and lift 5 Remove the probe with...

Страница 315: ...lease the button when the holder is inserted Push down firmly until the release button latches on 9 Fix the tube to the tubing clips 10 Reinstall the transfer head cover Press down until the release b...

Страница 316: ...sk 1 Select the maintenance action Clean Reagent Disk 2 Press The maintenance definition screen is displayed 3 Press 4 Open the main cover 5 Remove the reagent disk 6 Remove spillages and stains on th...

Страница 317: ...ish washer 12 Leave the sample tray to dry or dry it using tissues 13 Reinsert the sample tray Make sure to align the guides with the pins Press down firmly 14 Close the right service flap and the mai...

Страница 318: ...dex Bio instrument Operator time Approximately 10 minutes Prerequisites The system must be in Standby status Tools and materials required m Spare waste container m Purified water m Hypochlorite soluti...

Страница 319: ...To clean the waste container 1 Have the spare waste container ready The system periodically performs wash actions Therefore an external waste container must be connected at all times 2 Remove the tubi...

Страница 320: ...tem Maintenance actions 13 Wipe the waste inlet with a tissue 14 Reinsert the tubing adapter into the waste container Press down firmly 15 Press The maintenance action Prime Fluid System is automatica...

Страница 321: ...ex Bio instrument Operator time Approximately 10 minutes Prerequisites None Tools and materials required m Purified water m Hypochlorite solution 0 6 m Tissues m Water inlet filter a To replace the wa...

Страница 322: ...one o Place the retaining ring Press it firmly on the filter casing by placing your thumb across it an applying equal pressure all around 7 Reinsert the tubing adapter into the water container Press...

Страница 323: ...the main cover and the left service flap 9 Wipe all the inside of the flap with alcoholic cleansing tissues 10 Close the left service flap and the main cover 11 Push the transfer head to its home posi...

Страница 324: ...s The startup phase may take a few minutes When the instrument is ready for use the Overview tab is displayed and its status is Standby 19 Log on the system 20 Choose Utilities Maintenance Clean Housi...

Страница 325: ...rks slow for example when displaying information on the screen takes long or when it takes long until an action starts after you have initiated it Operator time Approximately 10 minutes Prerequisites...

Страница 326: ...equisites The system must be in Standby status Tools and materials required m Paper tissue m Photometer lamp assembly Lamp socket and cable are pre assembled m Cuvette segments m Clean cloth a To repl...

Страница 327: ...ssembly 11 Close the photometer lamp housing cover 12 Plug the power connector into its socket CAUTION Injury through hot parts Direct contact with hot parts may result in injury Do not touch the lamp...

Страница 328: ...the sample tray 14 Close the right service flap and the main cover 15 Press The system automatically initializes the instrument A message will inform you when the maintenance action is complete 16 Pr...

Страница 329: ...s a message when this maintenance action is due Operator time Approximately 2 minutes A service representative takes about 4 hours to perform preventive maintenance Prerequisites None a To initiate pr...

Страница 330: ...ettes If all cuvettes of this segment are empty you can reinsert it 5 Place a cuvette segment with empty cuvettes on the cuvette ring 6 Press Your are asked to insert a second cuvette segment with emp...

Страница 331: ...edex Bio System 8 General maintenance Maintenance actions 13 Press to accept the results 14 Remove the used cuvettes or replace them with empty ones If at this stage you press the results are not stor...

Страница 332: ...s must be closed Tools and materials required m Full external water container m Clean cloth a To initialize the degasser fluid sensor 1 Make sure that all covers on the instrument are closed 2 Choose...

Страница 333: ...re the filter A is not obstructed and air can get to it 8 Press to initialize the fluid sensor The system aspirates air through the tubing adapter and initializes the fluid sensor When initialization...

Страница 334: ...Remove the reagent disk o Close the main cover 2 Remove all cuvettes o Choose Overview o Open the main cover o Press a segment button o Remove the segment o Press to confirm the removal Remove all se...

Страница 335: ...General maintenance Maintenance actions 7 Open and remove the right service flap 8 Remove the sample tray Removing the rotor 9 Loosen the display fixation screw A 10 Tilt forward the display 11 Move t...

Страница 336: ...earing that you completely unscrewed 14 From the rotor motor slightly lift the retention pin move the motor away from the rotor and engage the pin to fix the motor in the removed position Make sure to...

Страница 337: ...d pull the spring loaded bearing away from the rotor position and with the other hand insert the rotor 20 Engage the right rotor bearings and fasten the fixation screws 21 Move the rotor to test its s...

Страница 338: ...e pins Press down firmly 27 Install and close the right service flap 28 Move the transfer head to its rightmost position 29 Install and close the left service flap and main cover Finishing the mainten...

Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...

Страница 340: ......

Страница 341: ...is chapter you will find information on messages generated by the Cedex Bio instrument and on how to use them and react to them About messages D 5 Message screen D 5 Acoustic signals D 6 Alarm monitor...

Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...

Страница 343: ...ways o Immediate feedback on user actions is displayed in a pop up message screen o Information concerning a problem that occurred during operation is reported as alarm messages in the alarm monitor...

Страница 344: ...can adjust the volume of all generated acoustic signals Utilities Configuration System Volume You can also turn on and off acoustic signals o Choose Utilities Configuration System Sound Effects to tu...

Страница 345: ...list to include in the list all alarm messages that are stored in the alarm database No color Off There are no unconfirmed alarm messages Yellow There is at least one unconfirmed alarm message You ne...

Страница 346: ...essage log on page D 9 6 When you have dealt with the issue press to confirm that you have dealt with the issue The message is deleted from the Not confirmed list but it remains in the other lists and...

Страница 347: ...ot confirmed list but they remain in the other lists and the alarm messages log which you can export When all messages are confirmed the Alarm LED is turned off not lit Exporting the alarm message log...

Страница 348: ...If the error recurs export the log files with the system remaining in this error condition If the error persists call service 0007 Failed to calculate linear regression Repeat the order If the error r...

Страница 349: ...instrument Remove the transfer head cover Check if the mixing mechanism is jamming or blocked Restart the system If the error persists call ser vice 5022 s Error temperature sensor analyzer heating sy...

Страница 350: ...y the Multi Slave control Restart the system If the error recurs export the log files with the system remaining in this error condition If the error persists call service 5061 d Air detected in the fl...

Страница 351: ...ed by the internal circuitry has been detected Restart the system If the error persists call service 6030 s Initialization of the rotor failed Switch off the instrument Remove the rotor and clean all...

Страница 352: ...ges and obstructions 7017 s The ISE sample sensor could not detect ISE Etcher Check the ISE tower and the ISE tubing for obstructions and leaks Ensure that the tubing is correctly inserted into the sa...

Страница 353: ...ISE sample sensor is not calibrated or no calibration data is available Perform Initialize ISE Unit If the error persists call ser vice 7058 s ISE sample sensor could not detect fluid Check the ISE to...

Страница 354: ...for obstructions and leaks Ensure that the electrodes are installed correctly If the error per sists call service 7075 s A timeout error occurred while performing an ISE measurement Restart the system...

Страница 355: ...Restart the system If the error persists call service 7115 d Dispensing into the ISE tower stopped because the ISE tower was not empty Check the ISE tower waste tubing Check the ISE tower outlet for o...

Страница 356: ...Restart the system If the error persists call ser vice 7134 s Motor X transfer overheated Switch off the instrument Check if the X transfer is blocked or jamming Ensure that the X transfer can be move...

Страница 357: ...he instrument Check if the rotor is blocked or jamming Restart the system If the error persists call ser vice 7150 d Initialization of the transfer failed due a level detection problem Switch off the...

Страница 358: ...trace file in the Host Settings in order to record HIF communication activities 12005 Host server could not send message to the host Select result and perform Retransmit If the problem recurs enable...

Страница 359: ...mber of standards used with number of calibrator cups and Last Is Water definition 20001 Replace Water Inlet Filter completed 20002 Clean Water and Waste completed 20003 Replace Photometer Lamp comple...

Страница 360: ...3 1 9 Messages and alarms Cedex Bio System List of alarm messages 20028 Daily Prepare Actions completed 20029 Database cleanup completed 20030 Initialize Degaser Fluid Sensor completed ID Message Com...

Страница 361: ...messages and the possible user actions About flags D 25 Safety D 27 List of flags D 28 D 28 V D 28 Ag Excess D 28 Cal D 29 Cal Error D 29 Calc Error D 29 Curv Dir D 30 Ep Unstab D 30 High Abs D 31 Hi...

Страница 362: ...Roche Diagnostics D 24 Operator s Manual Version 3 1 10 Result flags Cedex Bio System Table of contents Std Dev D 37 Test Rng D 38 Test Rng D 38...

Страница 363: ...tio results the flag with the highest priority of all constituent test results is displayed 2 Select the flagged result 3 Press to display all flags that were generated for this test A They are sorted...

Страница 364: ...r If one measurement does not generate the flag the counter is reset to zero Flag priority If the conditions in the Cedex Bio instrument are such that multiple flags were generated for a single measur...

Страница 365: ...ous material is spilled wipe it up immediately and apply disinfectant o If waste solution contacts your skin wash it off immediately with water and apply a dis infectant Consult a physician Personal i...

Страница 366: ...sult generated by re performing the test with the same running parameters repeated Message ID 120 Priority 32 Possible cause The user initiated the repeat Recommended actions No action required V Mean...

Страница 367: ...ailable Message ID 43 Priority 24 Possible cause Caused by an alarm that occurred during calibration because the calibration could not provide a usable result Recommended actions Check the flags of th...

Страница 368: ...ng End point unstable The absorbance values near the endpoint show unacceptable fluctuation Message ID 9 Priority 8 Recommended actions CAL 1 Repeat with fresh calibrators if only calibration is flagg...

Страница 369: ...control o If more than one control is flagged perform a calibration TEST o If a single sample result is flagged rerun with dilution o If the problem persists replace the reagent set High Act Meaning...

Страница 370: ...heck reaction plot 2 Repeat with fresh control TEST 1 Check reaction plot 2 Rerun with dilution or concentration if necessary Low Act Meaning Low activity This check is used to identify zero pipetting...

Страница 371: ...action plot 2 Repeat with fresh control TEST 1 Check reaction plot 2 Rerun with dilution or concentration if necessary Non Mono Meaning Curve not monotonic No calibration curve could be calculated bec...

Страница 372: ...ing A result where the quality control measurement is flagged Message ID 111 Priority 30 Possible cause The last QC measurement before this measurement was flagged The dependency is derived from the t...

Страница 373: ...the 2 standard deviation Message ID 29 Priority 26 Possible cause o The controls are not properly prepared o The controls are not properly positioned on the instrument Recommended actions 1 Repeat wi...

Страница 374: ...ring calibration the absorbance value of the lowest calibrator is below the lower defined range Message ID 12 Priority 7 Possible cause Deteriorated or outdated calibrator or reagent Recommended actio...

Страница 375: ...ired RR Meaning Below reference range The calculated concentration is less than the lower limit of the expected value range nnnnn is the actual value that was checked against Message ID 41 Priority 23...

Страница 376: ...ist of flags Test Rng Meaning PANIC value over upper Technical Limit The result is higher than the upper limit for the test Message ID 26 Priority 20 Test Rng Meaning PANIC value below lower Technical...

Страница 377: ...ation on how to deal with selected exceptional situations Introduction D 41 Dealing with exceptional situations D 42 Reacting to messages D 44 Detailed procedures D 46 Checking the printer status D 46...

Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...

Страница 379: ...controlled by performing all maintenance actions when they become due in the manner described in the Operator s Manual e See Maintenance actions on page C 8 Information about the status of the system...

Страница 380: ...tions given by the service representa tive e See Exporting diagnostics screens to USB on page D 51 The system does not work or does not switch on and the power switch light is off Instrument No electr...

Страница 381: ...me tests calibration does not work with special zero calibrators they need the Last Is Water definition on Possible remedies 1 Delete the current calibration order and redefine it 2 If the Last Is Wat...

Страница 382: ...or code column of the following table the most relevant information is the number the texts may be different on the actual instrument Message or error code Affected items Possible causes Comments 1 xx...

Страница 383: ...the Prime Fluid System mainte nance action e See Prime the fluid system on page C 11 3 If the problem persists contact your service representative 4 xxx 41 through 4 xxx 44 Various texts related to p...

Страница 384: ...er button is red 2 Scroll to display the Printer entry Clearing a paper jam a To clear the paper jam 1 Choose Stop Printing if required 2 Open the printer panel 3 Lift the printer paper roll from its...

Страница 385: ...paper and then automatically resumes printing 8 Choose Overview 9 The printer button should now be green and the status description should be OK Changing the mains fuses a To change the mains fuses 1...

Страница 386: ...lder 8 Remove the two fuses from the holder 9 Replace both fuses 10 Insert the fuse holder in the fuse box Press it in firmly until the clips engage 11 Connect the mains cable to the instrument 12 The...

Страница 387: ...until the holder disengages o Remove the holder with the fuse 3 Remove the fuse from its holder 4 Insert the new fuse in the holder 5 Insert the holder in the socket 6 Using the screwdriver lock the h...

Страница 388: ...an lift the reagent disk 2 Insert a pencil in the opening in the center of the underside of the reagent disk 3 Turn the reagent disk until the arrows of the reagent disk and the cover face each other...

Страница 389: ...d Lab Administrator or Administrator user rights to perform diagnostics actions and the system must be in Standby or Stopped status a To export the content of the screen 1 Press to start the diagnosti...

Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...

Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...

Страница 392: ......

Страница 393: ...unit You also will find a description of the main components and their principle operation Overview E 5 Abbreviations E 5 Measuring modes E 6 Principles of operation E 7 ISE solutions E 8 Hardware E 9...

Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...

Страница 395: ...mples are transferred using the same probe The ISE unit uses specific solutions some of which are kept in bottles on the ISE unit itself the others are transferred to sample tubes and placed on the sa...

Страница 396: ...diluted with system water 1 6 1 5 The dilution and mixing are performed automatically in the ISE tower Pipetting volumes Sample 15 L Diluent H2O 75 L ISE measurements ISE measurements and photometric...

Страница 397: ...urements are made 4 The sample is passed to the ion selective electrodes by the action of the peristaltic pump The exact positioning of segments is ensured by the ISE sample sensor 5 In the meantime t...

Страница 398: ...probe ion selective electrodes ISE tower and tubing during ISE maintenance Activator Sample area Activates the electrodes tubing and the ISE tower dur ing ISE maintenance It is also used for the init...

Страница 399: ...eft side panel Provides access to o ISE power supply o ISE connectors Do not remove this panel should be removed by your service representative only ISE cover Provides access to o Electrodes o ISE uni...

Страница 400: ...the aspiration action of the peristaltic pump Electrode block A block that contains the electrodes Power off e g power supply disconnected The ISE unit is OK and ready for use Blinking The fluid leve...

Страница 401: ...SE fluid bottles Two bottles are placed on the ISE unit o ISE Calibrator indirect urine bottle blue label o ISE Reference Solution bottle red label Level monitoring is performed on the basis of weight...

Страница 402: ...measurement and is performed automatically ISE Standby If there is no measurement for more than three minutes the ISE unit switches to Standby status In this status ISE Calibrator indirect urine and...

Страница 403: ...it is integrated in the maintenance of the main instrument The system software guides you through the maintenance procedures Maintenance actions are performed after a certain event or after a defined...

Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...

Страница 405: ...uide E 20 Starting the Prepare wizard E 21 Checking the status of the external fluid containers E 21 Checking the ISE status E 21 Performing maintenance actions E 22 Completing the preparation tasks E...

Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...

Страница 407: ...ntify the sample n a 2 Select the tests n a 3 Place the sample n a 4 Start the run 5 Monitoring the progress n a Overview 6 Validating results 1 View results n a Workplace Result Review 2 Handle flagg...

Страница 408: ...een n a 3 Start the QC measurement 4 Validate the QC results Workplace QC Status 5 Remove the controls 9 Finishing the shift 1 Check for unfinished orders Workplace Orders Choose Not Finished 2 Check...

Страница 409: ...it up immediately and apply disinfectant o If waste solution contacts your skin wash it off immediately with water and apply a dis infectant Consult a physician Personal injury or damage to the analy...

Страница 410: ...button If you refill the water empty the waste as well 2 Empty the waste container if required and confirm by pressing the waste button 3 Replace the cleaner bottle if required and con firm by pressin...

Страница 411: ...Follow the instructions given in Replacing electrodes on page E 32 When checking external fluid containers without using the Prepare wizard Choose Overview WARNING Injury through working solutions Di...

Страница 412: ...maintenance actions see Performing maintenance actions on page B 19 For details on how to perform individual ISE maintenance actions see ISE maintenance actions on page E 42 a To perform maintenance a...

Страница 413: ...that need calibrating Because the calibrations have already been performed as part of the Daily Prepare Actions maintenance action this step is not relevant for ISE operation 4 Press The Overview tab...

Страница 414: ...iew tab Interpreting the instrument status buttons A Relevant instrument status buttons Figure E 4 A Icon Meaning Possible action The ISE unit is in standby No action is required The expiration date o...

Страница 415: ...irst priori tized by color first priority being red followed by yellow and green and then according to the sequence in which they are listed below Analyzer main cover Press the button and on the Sys t...

Страница 416: ...on 1 and 2 The main calibration is typically performed during the Prepare phase and if there follows a second shift at the end of a shift It is integrated in the Daily Prepare Actions maintenance acti...

Страница 417: ...ration of the ISE Reference Solution o Crystallization of salts in the tubing which can cause blockages The ISE unit has a separate power supply Therefore the Standby status is maintained even if the...

Страница 418: ...d Shift but ton 3 Perform data backup 1 Press 2 Insert the USB stick 3 Press 4 Select the directory 5 Press 6 Press to proceed to the next stage in the End Shift wizard 4 Export support data 1 Press 2...

Страница 419: ...e the reagent disk Not relevant for ISE operation 9 Empty the waste container 1 Empty the waste container 2 Press the Waste button to confirm 3 Refill the water container 4 Press the Water button to c...

Страница 420: ...SE fluid bottles as soon as their expiration date has passed Their icon on the screen turns yellow For information on the stability of fluids see their package inserts WARNING Injury through working s...

Страница 421: ...w bottle 14 Insert the tubing adapter 15 Close the ISE cover 16 On the screen press to confirm the placement The system performs the Prime ISE Reference and Calib maintenance action 17 Press to close...

Страница 422: ...1 o ISE Solution 2 o ISE Sodium Electrode if you intend to replace this type of electrode o ISE Potassium Electrode if you intend to replace this type of electrode o ISE Reference Electrode if you int...

Страница 423: ...ange wizard The system rinses the electrodes with ISE Calibrator indirect urine to remove any sample residue and then drains all electrodes You are asked to remove the electrode Make sure to remove th...

Страница 424: ...e the electrode block lid 14 On the screen press the electrode button to confirm that you inserted the electrode If you inserted a dummy electrode press the button of the electrode you replaced with a...

Страница 425: ...area positions indicated on the screen 19 Press to confirm the placement and to start the action The system performs the Electrode Service maintenance action 20 Press to perform the main calibration...

Страница 426: ...materials required o Cotton swabs o ISE Deproteinizer o Small glass beaker a To clean the ISE tower Removing the ISE tower 1 Make sure the system is in Standby status 2 Shut down the system and switc...

Страница 427: ...ny visible clots 9 Pour ISE Deproteinizer in the beaker 10 Soak the ISE tower in the ISE Deproteinizer for about 5 minutes If there are clots in the tower you can extend the soaking time to 30 minutes...

Страница 428: ...Bio System Cleaning the ISE tower off the instrument Finishing the task 18 Switch on the instrument 19 Wait for initialization to finish 20 Log on to the system 21 Perform the maintenance action Clea...

Страница 429: ...tion E 41 ISE maintenance actions and their intervals E 41 ISE maintenance actions E 42 Safety information E 42 Daily prepare actions E 44 Activate electrodes E 45 Condition ISE tubing E 46 Initialize...

Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...

Страница 431: ...cs GmbH reserves the right to make any changes necessary without notice as part of ongoing product development Maintenance actions with no defined interval need to be performed in particular situation...

Страница 432: ...ng use contaminants may adhere to the probe As a result traces of analytes or reagents may be carried over to the next Make sure to perform the probe maintenance actions as soon as they are due in ord...

Страница 433: ...e o Follow standard laboratory practices especially when working with biohazard material Danger of poor measurement quality due to inadequate water quality Inadequate water quality may lead to incorre...

Страница 434: ...e system must be in Standby status Tools and materials required m ISE Deproteinizer m Activator m ISE Etcher m ISE Solution 1 m ISE Solution 2 a To perform the daily prepare actions 1 Prepare a tube w...

Страница 435: ...ator a To activate the electrodes 1 Select the maintenance action Activate Electrodes 2 Press The maintenance definition screen is displayed 3 Press to start the action A placement list is displayed 4...

Страница 436: ...ce action Condition ISE Tubing 2 Press The maintenance definition screen is displayed 3 Press to start the action A placement list is displayed 4 Place the Activator on the sample area position indica...

Страница 437: ...the ISE tubing You are asked to remove the tubing adapter from the ISE Reference Solution bottle 4 Open the ISE cover 5 Remove the tubing adapter from of the ISE Reference Solution bottle o Have a cle...

Страница 438: ...results Operator time Approximately 1 minute Prerequisites The system must be in Standby status Tools and materials required None Hazards and precautions None a To prime the ISE calibrator and ISE Re...

Страница 439: ...m electrode service 1 Select the maintenance action Electrode Service 2 Press The maintenance definition screen is displayed 3 Press to start the action A placement list is displayed 4 Place the Activ...

Страница 440: ...tomatically 1 Select the maintenance action Clean ISE Tower Automatically 2 Press The maintenance definition screen is displayed 3 Press to start the action A placement list is displayed 4 Place the A...

Страница 441: ...reen 5 Press to confirm the placement and to start the action The ISE tower is first washed The transfer head moves to the right side of the instrument You are asked to clean the ISE tower 6 Open the...

Страница 442: ...nsfer head moves to the right side of the instrument You are asked to clean the ISE tower 10 Open the main cover and then the left service flap 11 Clean the inside of the ISE tower with a cotton swab...

Страница 443: ...rials required m Activator a To initialize the ISE unit 1 Select the maintenance action Initialize ISE Unit 2 Press The maintenance definition screen is displayed 3 Press to start the action A placeme...

Страница 444: ...ed m Activator m Roche approved ISE pump tubing set a To replace the ISE pump tubing 1 Select the maintenance action Replace ISE Pump Tubing 2 Press The maintenance definition screen is displayed 3 Pr...

Страница 445: ...towards the back of the instrument 9 Disengage the tube connector plate 10 Remove the ISE pump tubing set Installing the tubing Install the tubing in the reverse order of removing it 11 Pull the new t...

Страница 446: ...the peristaltic pump lid The tube connector plate is automatically pushed into the tube connectors of the ISE unit 14 Close the ISE cover 15 Press to confirm completion of the action The system first...

Страница 447: ...ISE unit tubing 1 Select the maintenance action Replace ISE Unit Tubing 2 Press The maintenance definition screen is displayed 3 Press to start the action The placement screen is displayed 4 Place th...

Страница 448: ...but the process of removing the tube is the same o Have a tissue ready o Remove the tubing adapter by lifting it and place it on the clean lint free tissue o Remove the tube from its clips o Remove th...

Страница 449: ...w collector B by pulling it upwards and turning it from side to side as you do so Place the collector behind the wash station o Remove the locking ring C from the base of the ISE tower by turning it c...

Страница 450: ...E Reference Electrode and disconnect the tube leading to the ISE Reference Solution bottle 14 Disconnect the tubing connecting to the peristaltic pump Removing the tubing 15 Lift the tubing from its r...

Страница 451: ...the electrodes are properly mounted when closing the tension lever Close the electrode block cover 6 When connecting the tubing to the peristaltic pump connector block take care not to introduce shar...

Страница 452: ...sitions use the inner position o Insert the tubing adapter in the appropriate bottle 10 Close the ISE cover then the left service flap and the main cover 11 Press to confirm completion of the action C...

Страница 453: ...ubing adapter in the ISE Reference Solution bottle and to close the ISE cover 15 Insert the tubing adapter in the ISE Reference Solution bottle 16 Close the ISE cover 17 Press The system first pumps I...

Страница 454: ...ications on page B 139 3 Delete the ISE lot data e See Lot handling on page B 98 a To deactivate the ISE unit 1 Open the ISE cover 2 Remove the tubing adapters from the ISE solution bottles 3 Place th...

Страница 455: ...e connector plate 15 Remove all pinch valve caps and clamps store them in secure place e See Removing the pinch valve caps and clamps on page E 58 16 Close the ISE cover 17 Shut down the system and sw...

Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...

Страница 457: ...leshooting procedures and on how to deal with error messages Introduction E 69 Information about the ISE unit status E 69 Messages E 69 Safety E 70 List of ISE flags E 71 E 71 Air Fluid E 71 Air Iseca...

Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...

Страница 459: ...ther you need to intervene You can check the meaning of a button and its color using the Online Help o Buttons on the Overview tab lead to detailed information on the status of selected processes and...

Страница 460: ...cut which can lead to infection o If any biohazardous material is spilled wipe it up immediately and apply disinfectant o If waste solution contacts your skin wash it off immediately with water and a...

Страница 461: ...uid Meaning o The length of the sample segment was too short o The end of the sample segment was not detected It takes too long for a sample segment to reach the electrodes Message ID 19 Priority 4 Po...

Страница 462: ...if necessary o Repeat the measurement o Check the ISE tubing for leaks and blockages Cal Meaning A result where its calibration is flagged Message ID 110 Priority 29 Possible cause The calibration was...

Страница 463: ...ation of sample is too low for example dialyzed samples o The ISE Reference Solution tubing is blocked or leaking o The pH of the sample is too low for example pH 5 5 o One or more O rings are missing...

Страница 464: ...nd O Rings for proper mounting 4 Check the ISE tower and ISE tubing for leaks and blockages 5 Ensure that the ISE bottles on the ISE unit are available and check the level of fluid replace if necessar...

Страница 465: ...E Solutions 1 and 2 and check visually the correct flow of all ISE solutions and the ISE Reference Solution If successive calibrations are flagged 1 Repeat the calibration with fresh ISE Solutions 1 a...

Страница 466: ...standard deviation or below 3 standard deviation Message ID 35 Priority 27 Possible cause o Improper control is set o Proper control values mean value standard deviation are not specified Recommended...

Страница 467: ...rol material has been used 3 If the flag reappears check the calibration and reagent status Repeat with fresh reagent RR Meaning Above reference range The calculated concentration is greater than the...

Страница 468: ...as expired Recommended actions If only one calibration is flagged o Repeat the calibration If successive calibrations are flagged o Check whether the ISE Calibrator Indirect is in the correct position...

Страница 469: ...List of ISE flags Test Rng Meaning PANIC value over upper Technical Limit The result is higher than the upper limit for the test Message ID 26 Priority 20 Test Rng Meaning PANIC value below lower Tech...

Страница 470: ...ng on page E 57 o Check that the electrodes are mounted prop erly e See Replacing electrodes on page E 32 The cover of the peristaltic pump is open or not properly closed Close the cover The tubing ar...

Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...

Страница 472: ......

Страница 473: ...rial pipetted into any secondary cup alphanumeric sorting The listing of information in a printout or on a screen in a pre defined order by letters or numbers analyte The constituent in the sample tha...

Страница 474: ...liquid flow through the flow path before electrolyte measurement consumables A generic term for items that are used during test processing and must be replaced on a regular basis by the operator Exam...

Страница 475: ...button that allows access to the global software screens and that can be used at any time For example Start Stop or Alarm home position The position to which a certain part of the instrument returns o...

Страница 476: ...e analytical step performed by laboratory staff to reduce the analyte concentration in a sample mean The sum of values divided by the number of values measure point Time at which absorbance reading is...

Страница 477: ...The process of flushing a system for example the tubing with ISE Calibrator indirect urine and water to clear it of possible obstacles traces of fluid and of air bubbles profile A user defined set of...

Страница 478: ...be used for positive sample identification A sample tube contains sample of one specific specimen sample type sample type A type of sample that can be analyzed sampling stop An alarm level that indic...

Страница 479: ...turn around time 1 Time between receiving a sample and sending out the result unit A chosen reference quantity of an analyte used to compare quantities of the same dimension for example mol L g L or U...

Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...

Страница 481: ...es mapping B 160 applications B 137 activating B 139 calculation definitions B 155 calibration definitions B 153 control definitions B 154 deactivating B 139 definitions B 152 deleting B 144 general d...

Страница 482: ...e manually maintenance action C 12 Clean reagent disk and sample area maintenance action C 14 Clean rotor and heating channel maintenance action C 10 Clean the water and waste containers maintenance a...

Страница 483: ...se cleaning up C 23 exporting B 104 importing B 113 restoring B 113 date and time definitions B 167 deactivating ISE unit E 64 Default QC overview B 68 performing B 70 defining calibrator lots B 103 B...

Страница 484: ...les A 50 bottles for ISE E 11 checking external containers B 15 connecting external cleaner bottle B 128 connecting waste container B 128 connecting water container B 127 external connectors A 47 ISE...

Страница 485: ...tion E 26 maintenance E 39 measurement process E 7 measuring modes E 6 monitoring status E 24 operation E 17 overview E 5 preparing system E 20 priming calibrator E 48 principles of measurement E 7 pu...

Страница 486: ...Shift phase B 85 performing during Prepare phase B 19 Prime fluid system C 11 Prime ISE calibrator and reference E 48 Replace ISE pump tubing E 54 Replace ISE unit tubing E 57 Replace photometer lamp...

Страница 487: ...specifications A 66 photometer lamp replacing C 24 power interruption A 8 requirements A 65 precalibration A 29 Prepare phase short guide B 13 Prepare wizard B 15 preparing applications B 140 ISE uni...

Страница 488: ...screen A 90 reagents barcodes A 60 bottles A 53 cooling A 58 exchanging B 25 handling A 56 introduction to handling A 24 loading B 26 preparing B 22 B 25 removing B 25 replacing B 25 safety A 7 sets A...

Страница 489: ...when handling external bottles B 15 when handling reagents B 22 when removing reagent disk B 87 when troubleshooting ISE E 70 when using the barcode scanner B 89 with flags D 27 with maintenance actio...

Страница 490: ...stop option A 116 software importing B 113 updating B 113 software data handling specifications A 66 solutions ISE E 8 special characters A 76 specifications technical A 65 spillages safety A 9 stand...

Страница 491: ...4 user name B 11 U unfinished orders B 80 uninstalling applications B 143 units ISE E 9 laboratory B 156 transfer A 48 updating software B 113 upper case letters A 75 user interface overview A 20 A 69...

Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...

Страница 493: ...Revisions G...

Страница 494: ......

Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...

Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...

Отзывы: