Roche Diagnostics
D-10
Operator’s Manual · Version 3.1
9 Messages and alarms
Cedex Bio System
List of alarm messages
List of alarm messages
In the following table, the variables %s and %d are used, they represent text and numbers.
ID
Message
Comment
0000
Unknown message, id=%d!
0001
A fatal system exception software error (%d) occurred.
Restart the system. If the error recurs export the log files
while the system stays in the error condition. If the error
persists call service.
0002
An internal software error (%d) occurred while updating
the configuration.
Restart the system. If the error recurs export the log files
while the system stays in the error condition. If the error
persists call service.
0003
A software error occurred. Order (%d, %d) could not be
processed.
Delete the orders that are not processed yet. If the error
recurs export the log files while the system stays in the
error condition. If the error persists call service.
0004
A software error occurred. The unit is not ready, order
(%d) rejected.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0005
A mathematical exception arose.
Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0006
Last calculation cycle (%d) too short.
Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0007
Failed to calculate linear regression.
Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0008
The cycle range (%d..%d) is not valid.
Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0010
The reaction direction (%d) is invalid.
Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0011
An internal communication error occurred. The received
IPC message (%d, %d) could not be interpreted.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0012
A software error occurred. The status of an order is
invalid (%d.%d).
Delete and redefine the order. If the error recurs export
the log files with the system remaining in this error condi-
tion. If the error persists call service.
0017
The installed language version is out of date (%s, %s).
Install the language version that belongs to the currently
installed system software version. If the error persists call
service.
0018
The current system date and time are invalid.
Correct the date and time settings in Configuration. If the
error persists call service.
0019
Printer controller revision %d is not supported.
Call service.
Table D-1
List of alarm messages
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...