Roche Diagnostics
Operator’s Manual · Version 3.1
B-163
Cedex Bio System
7 Configuration
Configuration
Host definitions
Item
Values
Comment
Baud Rate
Data transmission rate for host interface communica-
tion.
Select one of the predefined values.
Handshake
Off
An exchange of signals between two devices when
communication begins in order to ensure synchroni-
zation.
Hardware
XON/XOFF
Line Mode
Number of bits and stop bits used during data trans-
mission.
Select one of the predefined values.
Parity Check
On
,
Off
An error detection technique that tests the integrity
of digital data in memory or on disk.
Communication
Mode
Off
Download to the
Cedex Bio
instrument is possible,
but upload to host is not.
Batch
Only manual upload to the host is possible.
Realtime
Results and queries are sent during processing.
Checksum
On
,
Off
A value used to ensure that data are stored or trans-
mitted without error. It is stored with the data. When
the data are retrieved from memory or received at the
other end of a network, a new checksum is computed
and matched against the existing checksum. A non-
match indicates an error.
Order Query
Mode
Query mode allows real-time communication during
order definition of regular, rerun, and repeated tests.
Off
No queries are sent.
Test orders have to be defined either manually or by
downloading them from the host.
Once
A query is sent only when the sample is identified the
first time.
Always
A query is sent every time the sample is identified.
Timeout
10 ... 300 s
Period of time the system will wait for a response
from the host when working in
Order Query Mode
.
This definition applies to routine, STAT and rerun
sample tests.
Send Result
Mode
Off
The results are sent to the host manually.
Complete
The results are automatically sent to the host when all
results of the order are available and accepted.
Immediate
Each result is automatically sent to the host as soon as
it has been accepted.
Send Raw Data
On
,
Off
Raw data include all measurement results made dur-
ing the measuring cycles, not just the calculated test
results.
Table B-21
Host definitions
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...