Roche Diagnostics
B-140
Operator’s Manual · Version 3.1
7 Configuration
Cedex Bio System
Applications
Preparing applications
Perform following steps for each new application. (Use the wizards to make the
definitions.)
e
For the general procedure of making changes to definitions, see
a
To prepare an application
1
Choose
Utilities
>
Applications
>
Laboratory Parameters
.
2
Expand the application entry.
Code
The Roche application code identifies the application. It is used for host
communication. If you work with a different code for a particular application, you
need to map the Roche application code to your own code.
e
Version
The version uniquely identifies a particular set of application definitions.
The version number increases whenever there is a change to the application
definitions, for example from version 1.0 to 1.1.
General definitions
3
Expand the
General
entry, select
Short Name
, and press
.
4
Change the short name (up to 5 alphanumeric characters) if required.
The short name affects the display of the test name in the user interface (for
example test buttons, or test names in the orders and results lists, and in all
printouts).
5
Press
and
change the long name (up to 30 alphanumeric characters) of the
test, if required.
The long name is a telling description of the test. It is particularly useful if the
short name is not generally familiar in the laboratory environment.
6
Press
and then
On
to activate the application.
If an application is not active, its associated tests are not available for use. You
cannot select them on the test selection screen. The application definitions remain
on the system.
7
Press
and select the buttons of the tabs the test should be available from.
e
For details, see
8
Press
to confirm the definitions.
Calibration definitions
For Roche calibrators, recommended definitions are automatically defined for each
application. You can change the sequence, interval, and the replicate value, if
required.
9
Expand the
Calibration
entry, select
Sequence
, and press
.
Roche recommend not to change the calibration definitions of Roche reagents.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...