8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface
155
Line Code 26: Dilution Factor
Example
26_••••2.1<LF>
q
Samples on the instrument can only be diluted. The
maximum dilution factor is 100.
22
Error in cuvette handling
23
Test aborted
24
Calculation not possible
25
No ISE Electrode
26
Activation not performed
27
ISE module not ready
28
Error in ISE module
29
Error in Abs. module
30
Error in Fp Module
31
Fatal HW Error
32
Activator not on board
33
Activator is empty
34
No activator definition found
35
Etcher not on board
36
Etcher is empty
37
No etcher definition found
38
Deproteinizer not on board
39
Deproteinizer is empty
40
No deproteinizer definition found
41
ISE module disabled
42
Fp module disabled
43
Cleaner not on board
44
Cleaner empty
45
No cleaner definition available
46
No ISE solution avaiable
47
No ISE solution definition avaiable
48
ISE solution is empty
49
Clot detected in sample
50
Cassette must be mixed first
Blocking status
flag code
Description
y
Blocking status flag codes
Description
Format
Range
Line Code
26
.
Dilution Factor
F7
1 to 100
y
Line Code 26: Dilution Factor