Appendix •
B
175
B
Appendix
B.1
Option: Wireless network (WLAN)
Preliminary note
Appendix B has been developed to explain the wireless
communication principles of the Accu-Chek Inform II
system
1
and to help your facility’s information technology/
management team(s) in effectively deploying the
Accu-Chek Inform II system on your wireless network.
Whether your meter has wireless capability or not
depends on the system configuration that you purchased.
Background
The Accu-Chek Inform II meter can only be configured
through a data management system to communicate wire-
lessly. The data management system is also used to set-up
and configure the meter to connect to one hospital specific
Wireless Local Area Network (WLAN
2
). WLANs use
electromagnetic waves in the 2.4 and/or 5.0 GHz frequency
range to wirelessly transmit data
3
. The Accu-Chek Inform II
system adheres to IEEE Standard 802.11g (2.4 GHz range)
4
.
The system is backwards compatible to 802.11b. During
wireless communication to an Access Point (AP), the
Accu-Chek Inform II meter recognizes the existent AP
WLAN protocol configuration (802.11b or 802.11g) and
automatically transmits data using the appropriate com-
munication protocol
5
.
1.
The Accu-Chek Inform II system is certified by the Wi-Fi
Alliance.
2.
WLAN is also commonly referred to as wireless LAN or
Wi-Fi.
3.
For the wireless functionality to work properly, the wire-
less module must first be configured by your system
administrator.
4.
While the Accu-Chek Inform II system adheres to the
802.11g standard, it uses only channels 1-11. Channels
12-14 are not used by the system.
5.
WLANs are organized in cells. A typical WLAN cell con-
sists of Access Point(s) that are connected to the (wired)
Local Area Network and one or more clients, e.g.,
Accu-Chek Inform II meters along with other clients such
as portable computers.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...