Safety reliably produced.
© RLS Wacon analytics GmbH ∙ +49 (0)5121 28126 0 ∙
Subject to
change without notice
53
Maintenance and service
Inserting a full reagent bottle
1.
Open the reagent bottle by twisting off the cap. Insert the
suction lance into the new bottle and screw the cap tight.
2.
Insert the full reagent bottle into the LIMES.
3.
Set the reagent level to 100 %. To do this, press the
"Reagent" and "Reset" buttons simultaneously and keep
them pressed until the LEDs "Reagent low" (blue) and
"Device fault" (red) light up simultaneously. For
more
information on the subject of "Reagent level" see page 51
4.
Now press and hold the "Reagent" button for 10 seconds to
pump reagent from the newly inserted bottle into the
measuring chamber and to press the air bubbles out of the
tube.
The reagent bottle is not included in the scope of delivery of the unit.
•
Use only original reagent type LHV in the 750 ml bottle.
•
Check the shelf life of the reagent used.
Danger of pollution
When handling the reagent, make sure that your eyes, skin and clothing do not come
into contact with the liquid.
•
Observe the instructions in the safety data sheets.
We accept no liability for permanent soiling caused by the dyes in the reagent or personal injury
resulting from improper handling of the reagent.
We recommend wearing suitable protective clothing when handling the reagent:
•
Workwear
•
Laboratory gloves
•
Eye protection / safety goggles
Irritant
Содержание 32-074165
Страница 76: ......