Supplies for Installation

Included with the LDCE

Leader cable

Meter leads

Available from RLE, sold separately

SeaHawk sensing cable

Available from Other Vendors

Multimeter with 

m

A capability

9V battery to power the LDCE (battery is NOT included with LDCE)

Please Note: 

The LDCE can only test for cable contamination or detected leaks.  It cannot 

test for a broken cable, and it cannot be used to verify that bulk cable has had 

connectors properly installed.

Install the 9V Battery

If you are using the LDCE for the first time, install the 

9 volt battery (not included with the LDCE) as follows:

1.  Screw in the two lock-down screws on the front 

cover using a 1/16” (1.59mm) allen wrench.

2.  Remove the two push button caps. Lift the cover 

and insert the battery.

3.  Loosen the lock-down screws of the LDCE until 

they are flush with the bottom cover.

4.  Replace the push-button caps for each screw you 

just loosened. 

Test the LDCE Battery Voltage

If you have used the LDCE before and have never replaced the 9-volt battery, 

test the battery before testing the sensing cable.  To test the battery:

1.  Connect the LDCE meter leads to your multimeter, as follows:

•  Connect the red lead to the Voltage (V) input.

•  Connect the black lead to the Common (COM) input.

Important!

 - Do not connect the leads 

to the wrong inputs, or you will blow the 

multimeter’s internal fuse.

2.  Set the multimeter to VDC.  If the 

multimeter is not auto-ranging, set it 

to the 20VDC scale.

3.  Push the System Check button on 

the LDCE and note the reading on 

the multimeter.  If it is less than 

8VDC, change the LDCE’s battery.

Test the SeaHawk Sensing Cable

The LDCE can test one cable or a string of cables, up to a total length of 5,000 

feet (1,524m).  To test the SeaHawk sensing cable for contamination:

1.  Power down the SeaHawk controller.  Disconnect the sensing cable to be 

tested and separate the white leader cable from the sensing cable.

2.  Connect the LDCE’s white leader cable to the sensing cable you wish to test.

SeaHawk LDCE Quick Start Guide

Thank you for purchasing a 

SeaHawk LDCE leak detection 

cable evaluator.  This guide 

outlines device installation and 

operation.

If you need further 

assistance, please contact RLE 

Technologies via our website 

- http://www.rletech.com/ (go 

to the Support link) or call us 

at 970-484-6510, Option 2.

v2.0

(07/2012)

To test the LDCE’s battery, connect it to the 

multimeter as shown.