- 34 -
WARNING
After the adjustment is completed, press MENU/ESC key to return to Detection Mode.
(If the detector remains in Maintenance Mode, it automatically returns to Detection Mode in ten
hours.)
7-2. Maintenance Mode
Mode
Item
LCD
display Details
Maintenance
Mode
Test Mode
2-0
2-0.0 Gas Test
2-0.1 Alarm Test
2-0.2 Fault Test
2-0.3 LED Test
2-0.4 Memory Test
Zero Adjustment
2-1
Perform a zero adjustment.
Span Adjustment
2-2
Perform a span adjustment.
Zero/Span Initialization 2-3
Initialize zero/span values.
Environmental Setting 2-4
2-4.0 Sensor Power Supply ON/OFF
2-4.1 INHIBIT Setting
2-4.2 Alarm Setpoint Setting
2-4.3 Alarm Delay Time Setting
2-4.4 Alarm Pattern Setting
2-4.5 Zero Suppression Pattern Setting
2-4.6 Zero Suppression Value Setting
2-4.7 Alarm Contact Specification Setting
2-4.8 Energized/De-energized Contact Setting
2-4.9 Zero Follower Selection
2-4.A Maintenance Mode External Output Setting
2-4.B External Output Adjustment
2-4.C Alarm Test External Output Setting
2-4.D Sensor Operation Start Setting
2-4.E Password Setting
2-4.F Sensor Fault Alarm Pattern Setting
2-4.H LOAD Voltage Adjustment
Display
2-5
Display various electrical settings.
This is not typically used by the user.
Switch to Factory
Mode
2-6
Not used.
Switch to User Mode
2-7
Returns to User Mode.