54
10
Connect the cable’s wires to the
sensor terminal strip. See the table
below for the wire colors. Type A
wire colors are shown in the image
to the right.
Note that the Type B cable has two
pairs of wires extending from the
cable. Tighten the zip tie to secure
the cable in place.
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Type A
Red
Green
White
Black
Type B
White wire of
white/yellow pair
Yellow wire of
white/yellow pair
White wire of
white/blue pair
Blue wire of
white/blue pair
11
Secure the top half of the main
unit’s housing to the bottom half.
12
Remove the sensor cover of the
new cable’s remote sensor
housing and install the sensor
board. The board will click into
place.
13
Connect the new cable’s plug into
the sensor board.
Secure the sensor cover of the
new cable’s remote sensor
housing.
Содержание RI-600
Страница 35: ...35 2 03 LCD Perform the LCD display test 2 04 Not used To 2 00 ...
Страница 43: ...43 2 4b P A Not used 2 4C ACAL Not used 2 4d BZZR Set the buzzer to ON OFF ...
Страница 44: ...44 2 4E AL R Set the gas alarm state contact to ON OFF 2 4F MA O Set the external output signal to ON OFF ...
Страница 45: ...45 2 4H LCD Set the LCD backlight to ON OFF No backlight by standard setting ...
Страница 60: ......
Страница 61: ......