5
2
Important Notices on Safety
2-1. Danger Statements
2-2. Warning Statements
DANGER
•
This is not an explosion-proof unit. Never attempt to measure a gas in an atmosphere over the
lower explosive limit.
WARNING
•
Before turning on the monitor, always check that the voltage is compliant with the specifications.
Operating on an unstable power supply may cause malfunctions.
•
Do not operate this monitor in a place where combustible/explosive gases or vapors are present.
Operating the monitor in such an environment will lead to extreme dangers.
•
Perform a span adjustment at fixed intervals.
•
Do not run the monitor’s power cable or remote sensor cable with power cables from another
device or with high-frequency or high-voltage cables.
•
If a high-frequency or high-voltage cable needs to intersect with the monitor’s power cable, it
should be orthogonally connected.
•
When wiring, be careful not to apply any stresses on the cables by pulling, tightening, twisting,
etc.
•
Do not disassemble/modify the monitor. It may invalidate the warranty of the performance.
Changing the settings without understanding them may cause alarm malfunctions. Please use
the monitor properly in accordance with the operating manual.
•
Do not use the monitor with it attached to a control device, equipment, etc.
WARNING
•
If an abnormality is found in the monitor, contact RKI immediately.
Содержание RI-600
Страница 35: ...35 2 03 LCD Perform the LCD display test 2 04 Not used To 2 00 ...
Страница 43: ...43 2 4b P A Not used 2 4C ACAL Not used 2 4d BZZR Set the buzzer to ON OFF ...
Страница 44: ...44 2 4E AL R Set the gas alarm state contact to ON OFF 2 4F MA O Set the external output signal to ON OFF ...
Страница 45: ...45 2 4H LCD Set the LCD backlight to ON OFF No backlight by standard setting ...
Страница 60: ......
Страница 61: ......