GasWatch 3 for CO and O
2
Appendix A: Maintenance Mode • 83
Performing a Gas Test
1. While in Maintenance Mode, press AIR to scroll to
GAS.TEST
.
2. Press and release POWER/MODE. The current gas reading displays. The bottom of the
LCD alternates between “GAS.TEST” and “BUZZ.OFF”.
3. Turn the regulator knob counterclockwise to open the regulator.
4. The instrument initiates alarm indications except for the buzzer. There is no buzzer
indication in the
GAS.TEST
menu even though the buzzer sounds in the event of a real
gas alarm condition.
5. Turn the regulator knob clockwise to close the regulator.
6. Unscrew the regulator from the calibration cylinder.
7. Remove the calibration cup from the GasWatch 3.
8. Store the calibration kit in a safe and convenient place.
9. Press and release POWER/MODE to return to the
GAS.TEST
item in Maintenance
Mode.
10. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.
Sensor/Battery Replacement Date (SEN.DATE)
The
SEN.DATE
item allows you to keep track of when the sensor and the battery were
replaced.
1. While in Maintenance Mode, press AIR to scroll to
SEN.DATE
.
2. Press and release POWER/MODE. The sensor replacement date displays.
GAS .TEST
GAS .TEST
20.9
%
0
ppm
BUZZ .OFF
20.9
%
0
ppm
SEN .DATE