24 • Chapter 3: Measuring Mode
GasWatch 3 for CO and O
2
2. The error code meanings are shown in the table below:
3. If the error code is anything but 031, the instrument cannot be used. Contact RKI
Instruments, Inc. as soon as possible.
If the error code is 031, press and release POWER/MODE to continue into Measuring
Mode if the instrument must be used temporarily.
CAUTION:
There is no datalogging function if you operate the instrument after a 031
system failure. Contact RKI Instruments, Inc. as soon as possible.
Data Logging
The GasWatch 3 logs Measuring Mode gas readings, alarm data, and calibration data to its
internal memory. Logged data can be download it to a computer via the infrared
communications port on the front of the unit.
The data logging capacity depends on how often the GasWatch 3 stores data and how often the
GasWatch 3 is turned on and off. The table below illustrates how much data logging time is
available for the various interval times. It assumes that the unit is only turned on once and
there are no alarms. The data logging interval time must be set using the GasWatch 3
Datalogging Program.
Table 5: Error Code Explanation
Error Code
Explanation
000
ROM failure
010
RAM failure
020
FRAM failure
031
FLASH memory failure
082
Temperature sensor failure
Table 6: Data Logging Capacity
Interval Time
Data Logging Capacity
10 seconds
10 hours
20 seconds
20 hours
30 seconds
30 hours
1 minute
60 hours
3 minutes
180 hours
5 minutes
300 hours
10 minutes
600 hours