55
•
Maintenance
AirLink 6940XP Operator’s Manual
Replacing the Desiccant
WARNING: Do not remove the sensor housing cap or enclosure lid while the circuits are
energized unless the area is determined to be non-hazardous. Keep the sensor
housing cap and enclosure lid tightly closed during operation.
Each AirLink 6940XP comes with a desiccant bag installed in the junction box. The contents are
blue when it is dry. As the desiccant absorbs moisture, it turns amber. Periodically check the
desiccant and replace it if it has turned amber.
Replacing the Battery
Under normal conditions, the AirLink 6940XP’s battery pack lasts 6 months (45 days if a PID
sensor is installed). When the battery indicator is blank, approximately 2 weeks of battery life
remain.
The presence of gas will increase the radio transmission rate and decrease the battery lifetime. It is
recommended that the battery pack be replaced when the voltage reported is 3.0 volts or less. See
page 29 to determine the battery voltage.
WARNING: Do not remove the sensor housing cap or enclosure lid while the circuits are
energized unless the area is determined to be non-hazardous. Keep the sensor
housing cap and enclosure lid tightly closed during operation.
WARNING: Only use RKI-supplied replacement battery packs.
CAUTION:
The internal components can be static sensitive. Use caution when opening the
enclosure and handling internal components. DO NOT use any metal objects or
tools to remove the batteries.
1. Press and hold
SUB
for approximately 6 seconds until “OFF” shows on the display screen.
2. Unscrew the enclosure lid, and set it aside.
3. Grip the front panel handles, lift the internal system out of the enclosure, and rest it against the
edge of the enclosure.
CAUTION:
Be sure power to the AirLink 6940XP is off before pulling out the internal system.
NOTE:
Disconnecting the sensor connector plug from the sensor housing connector
and the antenna fitting will allow for the complete removal of the internal
system from the device enclosure. Disconnecting the internal system may
provide ease in accessing the control board terminals for wiring. Reconnect
the sensor connector plug before reinstalling the internal system.