51
•
Sensor Settings and Calibration
AirLink 6940XP Operator’s Manual
Reset Zero & Calibration Values
Resetting the zero and calibration settings of the sensor element will allow the currently stored
zero and calibration values to be rest without having to reconfigure all of the other operational
settings like with the Return to Factory Defaults option.
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
MENU
for approximately 6 seconds.
Press and release
ADD
or
SUB
to navigate to the desired sensor settings menu, then press
MENU
to select the menu.
2. Press
ADD
until the “Return Zero/Cal to Default” menu item appears. Press
MENU
to select
the “Return Zero/Cal to Default” menu item.
3. Press
ADD
to select “Yes” to advance to the “Reset Zero & Cal Only?” confirmation screen.
To return to the Sensor Settings menu, press
SUB
to select “No”. Pressing
MENU
or
BACK
also returns the instrument to the Sensor Settings menu.
4. If “Yes” is selected, the following confirmation screen appears. Press
ADD
to reset the Zero
and Calibration values.
To return to the Sensor Settings menu without reseting the Zero and Calibration Values,
press
SUB
. Pressing
MENU
or
BACK
also returns the instrument to the Sensor Settings
menu.