AirLink 6940XP Operator’s Manual
Sensor Settings and Calibration
•
50
Zero/Calibration Timer Information
The zero/calibration time information screen displays the following information:
• Days since the sensor assembly was last zeroed
• Days since the sensor assembly was last calibrated
This screen is for informational purposes only.
NOTE:
The Days Since Last CAL field automatically updates after an Auto Cal. When
performing a Manual Cal, the gas reading must be adjusted by at least one button press
to get this field to update.
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
MENU
for approximately 6 seconds.
Press and release
ADD
or
SUB
to navigate to the desired sensor settings menu, then press
MENU
to select the menu.
2. Press and release
ADD
until the “Last Time Zero/Cal” menu item appears. Press
MENU
to
select the “Last Time Zero/Cal” menu item.
3. Review the information displayed.
4. Press
MENU
or
BACK
to return to the Sensor Settings menu.