62 • VOC Pro Transmitter
Calibrating the Sensor (Auto Cal)
You should ONLY calibrate the sensor after completing the zero process.
1. Press the ADD button to select “Yes” to begin the calibration process and to advance to
the calibration confirmation screen. If you do not wish to calibrate the sensor, press the
SUB button to select “No” to advance to the sensor radio address setting screen.
2. Press the ADD button to select “Yes” to confirm that you want to calibrate the sensor and
to continue to the concentration setting screen. If you do not wish to continue to calibrate
the sensor, press the SUB button to select “No” to advance to the sensor radio address
setting screen.
3. Use the ADD and SUB buttons to adjust the concentration to the calibration gas being
used.
NOTE:
For 10.6 eV and 11.7 eV PID sensors, if you are calibrating with a gas other than
isobutylene, you must take the calibration gas’ factor to isobutylene into account
when setting the calibration gas concentration.
4. Press the MENU button to save the gas concentration setting and to advance to the
calibration start screen.
5. Unscrew and remove the rain guard from the assembly.
6. Install the calibration cup to the VOC Pro’s sensor housing.
7. Screw the regulator into the calibration cylinder.
8. Use the sample tubing to connect the regulator to the calibration cup.