30
•
Calibration, CO Sensor
Calibration, CO Sensor
This section describes how to calibrate the CO sensor in the sample-draw detector. It includes
procedures to prepare for calibration, set the zero reading, set the response reading, and return
to normal operation.
NOTE:
Calibrating the sample draw detector may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before continuing.
NOTE:
This procedure describes calibration using a demand flow regulator.
Preparing for Calibration
NOTE:
If you can verify a fresh air environment, it is not necessary to use a zero air
calibrating sample to set the zero reading at the controller.
1. Follow the instructions in the controller’s operator’s manual for entering calibration mode.
2. Screw the regulator into the zero air calibration cylinder.
3. Open the housing door.
4. Set a voltmeter to measure in the millivolt (mV) range.
5. Connect the voltmeter to the test points on the CO amplifier.
Plug the positive lead into the red (+) amplifier test point; plug the negative lead into the
black (-) amplifier test point.
6. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 ppm and a fullscale setting of 300 ppm, the correct
output is 167 mV.
167 (mV) = (50/300) X 400 +100
Setting the Zero Reading
1. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s
inlet fitting.
This step is not necessary if you verified a fresh air environment earlier
in this procedure.
2. Allow the sample draw detector to draw sample for one minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).