www.rkiinstruments.com
35-3001-NS-03
Sample Draw Pump
Operator’s Manual
Part Number: 71-0472
Revision: P1
Released: 11/16/18
Страница 1: ...www rkiinstruments com 35 3001 NS 03 Sample Draw Pump Operator s Manual Part Number 71 0472 Revision P1 Released 11 16 18...
Страница 2: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...
Страница 3: ...ng the Sample Draw Pump 8 Connecting the Sample Lines to the Sample Draw Pump 9 Wiring the Sample Draw Pump to a Power Source 10 Start Up 12 Introducing Incoming Power 12 Low Flow Contact Operation 13...
Страница 4: ...the detector head manual for information specific to the detector head WARNING When using the 35 3001 NS 03 you must follow the instructions and warnings in this manual to assure proper and safe opera...
Страница 5: ...weather and corrosion resistant It is suitable for installation where general purpose equipment is in use Pump Connector Pump _ O X Y G E N B W Flow Baffle OXY S AM P 2 AM P S Sensor Flow Control Valv...
Страница 6: ...housing by using conduit or an appropriate cable bushing Internal Components This section describes the sample draw pump s internal components see Figure 1 Figure 2 illustrates how the gas sample move...
Страница 7: ...flow control valve adjusts the flow rate to the detector Turn the valve s knob counterclockwise to increase the flow and clockwise to decrease the flow Status LEDs Two status LEDs are above the flowm...
Страница 8: ...r source Mounting the Sample Draw Pump 1 Select the mounting site Consider the following when you select the mounting site Is there enough room to open the housing door and make wiring and tubing conn...
Страница 9: ...using to the mounting surface Connecting the Sample Lines to the Sample Draw Pump Figure 4 Fitting Arrangement Installing the Inlet Line 1 Loosen the nut on the inlet fitting until 3 threads are visib...
Страница 10: ...hs based on tubing size CAUTION If you use flexible sample tubing polyurethane is acceptable use an appropriate insert to seal the connection between the tubing and the inlet fitting 3 Hand tighten th...
Страница 11: ...hown in Figure 5 Figure 5 Wiring the Sample Draw Pump to a Power Source AMP 1 LO W F LO W AMP 2 LO W F LO W AMP 1 OXY AMP 2 OR OXY 24 VDC DC Ground Not Used Resistive Load Device Contact Monitoring De...
Страница 12: ...f 2 0 SCFH still provides good sensor response at the flow through adapter RKI Instruments Inc does not recommend a flow rate of lower than 2 0 SCFH Turn the valve s knob counterclockwise to increase...
Страница 13: ...a flow rate between 2 0 and 3 0 SCFH See Specifications on page 4 for more discussion of an appropriate flow rate If necessary use the flow control valve to adjust the flow rate Turn the valve s knob...
Страница 14: ...h end of the particle filter from the flow system tubing 3 Remove the old particle filter 4 Install the new particle filter Be sure the arrow on the particle filter is pointing away from the tubing co...
Страница 15: ...ppropriate flow rate 6 Make sure the sample draw pump s Fail LED is off Parts List Table 4 lists replacement parts and accessories for the sample draw pump Table 4 Parts List Part Number Description 0...