RJG Sensor PreCheck Скачать руководство пользователя страница 25

RJG, INC. SENSOR PRECHECK | USER TOOLS

19

2

3

5

5

4

4

1

6

MANUAL SENSOR 

FORCE TESTS

STRAIN GAGE SENSORS

After running an 

automated test, 

select

   

one of the following: 

• 

1

 

Test Basic and Device    

Force 

• 

2

 

Touch to Begin Force 

Test

• 

3

 

Force Test All

NOTE 

The 

application will time 

out if no signal is 

received from the 

sensor in a specified 

length of time; be 

ready to apply force 

to the sensor(s).

Select

   

4

 

End Force 

Test

 

after the 

5

 

Signal 

Detected 

screen is 

displayed. 
The 

6

 

No Signal 

Detected

 

screen will 

be displayed if the 

test was unsuccessful. 

Refer to 

PAGE 30

 for 

troubleshooting.

Содержание Sensor PreCheck

Страница 1: ...USER GUIDE Testing and diagnostics Error Reporting Sensor Validation Training and Technology for Injection Molders...

Страница 2: ...PRINT DATE 01 19 2021 REVISION NO 2 4 2...

Страница 3: ...ON 2 RUN AN AUTOMATED SENSOR TEST 3 RUN A MANUAL SENSOR FORCE TEST 4 STRAIN GAGE SENSORS 4 PIEZOELECTRIC SENSORS 5 GENERATE AND DISTRIBUTE REPORTS 6 OVERVIEW 7 SENSOR PRECHECK HARDWARE 8 AP 32C 8 LYNX...

Страница 4: ...AND INSTALL FROM TABLET 24 REFRESH APPLICATION 24 REFRESH SENSOR PRECHECK LIST 25 REFRESH SENSOR LIST 26 UPDATES 27 APPLICATION 27 FIRMWARE 27 TROUBLESHOOTING 29 STRAIN GAGE SENSORS 30 COMMUNICATION...

Страница 5: ...it does not guarantee that the same results as those described herein will be obtained Nor does RJG Inc guarantee the effectiveness or safety of any possible or suggested design for articles of manuf...

Страница 6: ...Sensors Basic Force Test Specifications Compatible Sensor Models Strain Gage LS B 127 50 125 500 2000 LS B 159 4000 LES B 127 50 125 500 2000 LES B 159 4000 MCSG B 60 50 250 MCSG B 127 125 500 2000 MC...

Страница 7: ...CHECK QUICK START GUIDE START EQUIPMENT AND APPLICATION 2 RUN AN AUTOMATED SENSOR TEST 3 RUN A MANUAL SENSOR FORCE TEST 4 STRAIN GAGE SENSORS 4 PIEZOELECTRIC SENSORS 5 GENERATE AND DISTRIBUTE REPORTS...

Страница 8: ...32C is off Connect the Lynx cable to the AP 32C 3 Lynx input and the sensor s to be tested ensure the provided USB WiFi adapter is inserted in the AP 32C USB port Select the RJG Sensor PreCheck 4 app...

Страница 9: ...nt model s Select the appropriate response for the connected sensor s to continue Enter the 2 Mold Number in the provided field Select 3 Begin Automated Test to test the sensor s The 4 Progress Bar wi...

Страница 10: ...E The application will time out if no signal is received from the sensor in a specified length of time be ready to apply force to the sensor s STRAIN GAGE SENSORS After completing an automated test se...

Страница 11: ...adapter cable during multi channel piezoelectric sensor testing will create false test results After completing an automated test select 1 Test Basic and Device Force OR 2 Touch to Begin Force Test to...

Страница 12: ...ted sensors Select the 1 share button located at the top right hand corner of the 2 completed test screen The 3 report screen will display Select the 4 email function from the 3 report screen Select 5...

Страница 13: ...ARDWARE 8 AP 32C 8 LYNX SENSOR CABLE 9 TABLET 9 SENSOR PRECHECK APPLICATION 10 AUTOMATED SENSOR TEST 11 MANUAL STRAIN GAGE SENSOR FORCE TEST 12 MANUAL PIEZOELECTRIC SENSOR FORCE TEST 13 SENSOR TEST RE...

Страница 14: ...CK 8 SENSOR PRECHECK HARDWARE AP 32C The 1 AP 32C verifies proper operation of Lynx cavity pressure sensors and includes the following 2 Lynx Connector 3 Power Supply Connector 4 Power Supply Indicato...

Страница 15: ...or testing TABLET The tablet provided with the AP 32C is preloaded with the Sensor PreCheck application and provides a user interface for testing sensors and generating sensor test reports Refer to th...

Страница 16: ...reCheck application tests labels and generates a report for up to 30 sensors simultaneously The application provides the following tests 1 Automated Sensor Tests Multiple or Single Sensor Manual 2 Str...

Страница 17: ...EST The automated sensor test will test each sensor connected to the Sensor PreCheck simultaneously for the following 3 Zero Shift for Strain Gage only 4 Drift for Piezoelectric Sensors only 1 Communi...

Страница 18: ...er for the Sensor PreCheck to evaluate if the sensor is detecting force NOTE The application will time out if no signal is received from the sensor within a specified length of time be ready to apply...

Страница 19: ...out if no signal is received from the sensor within a specified length of time be ready to apply force to the sensor s The manual piezoelectric sensor force test will display a 2 test in progress pag...

Страница 20: ...is outside normal but not in an Alert range The 6 single sensor test results page indicates if the tested sensor has passed communication gage wire zero shift strain gage only drift piezoelectric only...

Страница 21: ...share button to view a sensor test report after completing sensor testing Reports detail 2 incomplete tests 3 alerts 4 warnings and the 5 number of sensors that passed testing The reports also show i...

Страница 22: ...S 2 F S PASS FAIL LES B 159 4000 PASS FAIL 5 F S 2 F S 2 F S PASS FAIL MCSG B 60 50 250 PASS FAIL 20 F S 15 F S 15 F S PASS FAIL MCSG B 127 125 500 2000 PASS FAIL 9 F S 6 F S 6 F S PASS FAIL MCSG B 1...

Страница 23: ...17 RJG INC SENSOR PRECHECK USER TOOLS USER TOOLS AUTOMATED SENSOR TESTS 18 MULTIPLE SENSORS 18 MANUAL SENSOR FORCE TESTS 19 STRAIN GAGE SENSORS 19 PIEZOELECTRIC SENSORS 20 SENSOR TEST REPORTS 21...

Страница 24: ...er selection If desired enter the mold number in the 2 Mold Number field Select 3 Begin Automated Test to test all sensors The 4 Progress Bar will indicate the test progress Wait for the test to compl...

Страница 25: ...Force 2 Touch to Begin Force Test 3 Force Test All NOTE The application will time out if no signal is received from the sensor in a specified length of time be ready to apply force to the sensor s Sel...

Страница 26: ...llowing 1 Test Basic and Device Force 2 Touch to Begin Force Test 3 Force Test All NOTE The appli cation will time out if no signal is received from the sensor in a specified length of time be ready t...

Страница 27: ...lect either 2 NO to remain connected or YES to disconnect from the Sensor PreCheck and email the report NOTE The email button must be selected to generate the report if the email button is not selecte...

Страница 28: ...USER TOOLS RJG INC SENSOR PRECHECK 22...

Страница 29: ...PRECHECK SETTINGS SETTINGS INSTALLATION AND SETUP 24 APPLICATION DOWNLOAD AND INSTALL FROM TABLET 24 REFRESH APPLICATION 24 REFRESH SENSOR PRECHECK LIST 25 REFRESH SENSOR LIST 26 UPDATES 27 APPLICATI...

Страница 30: ...Applications icon from the tablet home page Select the 2 Play Store from the list of Apps Enter 3 Sensor PreCheck in the search bar Select the 4 Sensor PreCheck application Select 5 Install from the...

Страница 31: ...ted Sensor PreCheck does not appear on the list of Sensor PreChecks swipe the tablet screen in a downward motion to 1 refresh the application for any recently added or removed Sensor PreCheck systems...

Страница 32: ...nsors appear upon application start up or if a recently connected sensor does not appear on the list of sensors select 1 Retest to refresh the application for any recently added or removed sensor conn...

Страница 33: ...ET ON PAGE 24 FIRMWARE Firmware updates for the Sensor PreCheck are installed from the tablet Open and connect to the Sensor PreCheck Select the 2 UPDATE FIRMWARE button Any firmware updates will be p...

Страница 34: ...SETTINGS RJG INC SENSOR PRECHECK 28...

Страница 35: ...RS 30 COMMUNICATION FAILURE 30 GAGE WIRE TEST FAILURE 30 ZERO SHIFT FAILURE 30 FORCE TEST FAILURE 30 PIEZOELECTRIC SENSORS 31 COMMUNICATION FAILURE 31 FORCE TEST FAILURE 31 FORCE TEST PASS WITHOUT APP...

Страница 36: ...test check the sensor pocket installation dimensions If the sensor fails again request an RMA for return from RJG ZERO SHIFT FAILURE Remove sensor from tool and retest Failure could be caused by prelo...

Страница 37: ...the sensor from the tool and retest the sensor If the sensor passes check that the pin moves freely The pin should slide forward and backward easily The pin should also spin easily If the sensor fails...

Страница 38: ...roved cleaner Follow instructions at https rjginc com paperclip product_downloads 547 cleaning connectors_cables pdf Retest after cleaning If the test fails after cleaning request an RMA from RJG 2 Si...

Страница 39: ...inc com know how knowledge base RJG s searchable virtual help library Topics include Machine Interface eDART Data Manager Networking Sensor Communication Extracting eDART Data Advanced System Overview...

Страница 40: ...UBLESHOOTING RJG INC SENSOR PRECHECK 34 CUSTOMER SUPPORT Contact RJG s Customer Support team by phone or email RJG Inc Customer Support P 800 472 0566 Toll Free P 1 231 933 8170 https rjginc com suppo...

Страница 41: ......

Страница 42: ...S E A PTE LTD Singapore Republic of Singapore P 65 6846 1518 sales sg rjginc com en rjginc com FRANCE RJG FRANCE Arnithod France P 33 384 442 992 sales fr rjginc com fr rjginc com CHINA RJG CHINA Chen...

Отзывы: