Operating Manual Version 6.3-04 of 10.10.18
34
ANNEX 1
Guidance and manufacturer's declaration –
electromagnetic compatibility (EMC)
Intended healthcare environments
–
Professional and Home
Emissions
Classification
—
Standard
EN 55011
(idt CISPR 11)
Class A or B
B
—
Group 1 or 2
1
—
Conducted RF Emissions
N/A
NOTE 1
Radiated RF Emissions
PASS
Disturbance Power (if applicable)
N/A
NOTE 2
Harmonic Distortion per EN 61000-3-2 (Class A,
B, C, D)
N/A
NOTE 1
Voltage Fluctuations and Flicker per EN 61000-3-3
N/A
NOTE 1
Immunity
Electrostatic Discharges EN 61000-4-2
PASS
Radiated RF EM Fields and Proximity Wireless
fields EN 61000-4-3
PASS
Electrical Fast Transients and bursts EN 61000-4-4
N/A
NOTE 3,4
Surges EN 61000-4-5
N/A
NOTE 3,5
Conducted Disturbances, induced by RF fields
EN 61000-4-6
N/A
NOTE 3,4
Voltage Dips and Interruptions EN 61000-4-11
N/A
NOTE 1
Rated Power-frequency Magnetic Field EN
61000-4-8
PASS
Supplementary information:
NOTE 1) EUT is powered by internal battery 9V DC.
NOTE 2) If applicable Radiated RF Emissions, Disturbance Power is not necessary.
NOTE 3) The test is applicable to all d.c. power PORTS intended to be connected
permanently to cables longer than 3 m.
NOTE 4) SIP/SOPs whose maximum cable length is less than 3m in length are
excluded.
NOTE 5) This test applies only to output lines intended to connect directly to
outdoor cables.