Continue table 4
Fault
Possible cause
Troubleshooting method
There is no energy on
the add-on electrode,
while the energy on the
built-in electrode is felt.
The add-on elec-
trode malfunc-
tion.
There is no con-
tact between the
plug and jack.
The incompati-
ble add-on elec-
trode.
Replace the electrode.
Check the connection between
the add-on electrode plug and
the SCENAR device jack.
Use only the add-on electrodes
supplied by the manufacturer
of the SCENAR device.
There is no energy both
on the add-on and built-
in electrodes when the
add-on electrode is con-
nected, but the energy on
the built-in electrode is
felt when the add-on
electrode is not connect-
ed.
Short circuit in
the add-on elec-
trode cable or
plug.
The incompati-
ble add-on elec-
trode.
Replace the electrode.
Use only the add-on electrodes
supplied by the manufacturer
of the SCENAR device.
The SCENAR device
turns off when there is a
skin contact.
Bad skin contact
or the skin is very
dry.
Check the settings of the
AOff
. Refer to item 6.3.1.
8.2 Other malfunctions shall be repaired only by the manufacturer.