Operating Manual Version 7.3-03 of 28.07.20
27
6.3.1 Setting the Automatic Turn-Off Time (
AOff
)
The
AOff
menu item enables or disables automatic turn-off of
SCENAR.
When the
AOff
is set to
Off
, the automatic turn-off is disabled.
(At-
tention! This may cause faster battery exhaustion when you leave the
device switched on for a long time)
.
When the
AOff
is set to
On
, then when no buttons are pressed and
no contact with the skin has been detected within 30 sec, SCENAR will
switch to the standby mode (see p.6.1.9).
If
AOff
is set to
5m
(
10m
or
20m
)
, then when no buttons pressed
and no contact with the skin has been detected within 30 sec, SCENAR
will also switch to the standby mode. Moreover, in the
Dose 0
mode
SCENAR will switch to the standby mode in 5 (10 or 20) minutes even if
there was a contact with the skin. This mode is intended to treat one point
for a specified time.
Attention!
In the
‘
AOff Off
’ and ‘
AOff 5m
’
(
10m
or
20m
) modes,
taking the electrode up from the skin as well as pressing any
button resets the timer.
The differences of the automatic turn-off modes are given in the
Table 3.
Table 3
Automatic turn-off time when no buttons pressed
AOff
No contact
Contact
Timer reset
Off
never
never
never (only when buttons
pressed)
On
30 sec
never
Every time the electrode
contacts / is taken up
from the skin
5m
5 min
10m
10 min
20m
20 min
6.3.2 Setting the Backlight Time
(Lght)
The
Lght
menu item is the time of screen backlighting for the fol-
lowing events – pressing any button, reaching the
Dose
or
Zero,
expi-
ration of the first second of measuring in the
Dose state
or expiration
of every 30 seconds.