6. Number of Interruption :
This Screen Displays the total no. of times the Axillary
Supply was Interrupted. Even if the Auxiliary supply is
To reset No of Interruption manually see section
Resetting Parameter 3.2.3.1
interrupted count will be maintained in internal memory.
VYB
VRY
KVAr
Sys
V
A
KVA
Min
KW
VBR
Angle
Max
x1000
KVArh
x1000
x1000
IN
Ph R
Hz
KVAh
Ph Y
P.F.
KWh
Ph B
KAh
%THD
Demand
7. Analog Output ( optional ) :
This module provides two d.c. isolated outputs .There are two output options
1) Two 0 - 1mA outputs , internally powered .
2) Two 4 - 20mA outputs , internally powered .
The 0 -1mA output module has an 0V return on each end of the 4 way connector
( Please refer section 15 for connection details )
On both modules the output signals are present on pins A1(Analog Output 1) &
A2 (Analog Output 2)
These outputs can be individually assigned to represent any one of the measured and
displayed Parameters.
All settlings are user congurable via the user interface screen. See Analog o/p selection
( section 3.2.4.3 & section 3.2.4.4 ) for details .
* Note : Refer diagrams 1 & 2