User’s Manual
12 of26
long distant transportation.
3. Assembly of its components and ensure that all connections are firmly in place.
4. Fill the water tank with distilled water only.
5. Installation treatment handpiece and connect the power supply
6. Switch on the device and test all functions / parameters of the system
3.1 Equipment Detailed List
The following is the detailed list of equipment of this device:
:
Main host
The laser treatment handpiece 3pcs (1064nm & 532nm& black doll)
Handpiece support(assembled in device
)
Protective glasses
Protective goggles
Power cable
Funnel
Plastics hose
User’s manual
3.2 Installation Requirements
Before unpacking the Laser treatment systems, ensure that working environment conforms to
the requirements of this section.
Before assembly of the device’s components, ensure that high humidity is cleared away by
opening the packing and staying the device for one day.
3.2.1 Location Requirements
The device should be positioned in an area with adequate ventilation away from devices that
produce heat (i.e. refrigerator).
3.2.2 Electrical Requirements
Before the Laser leaves the factory, it has already been marked the local rated voltage
according to the customers’ request. It corresponds to the electricity requirement as follows
:
AC230V±10
%,
10A
,
50/60Hz. or AC115V±10
%,
20A
,
50/60Hz
Input electric current can’t have momentary change, electric voltage or electric current peak
、
sink.
It is recommended that this device use a dedicated power supply with its own circuit breaker.
Ensure that the electric voltage rating of this instrument (AC230V or AC115V refer to
system labeling) matches the electric voltage of the outlet.
3.2.3 Environment Requirements
Air quality
:
No caustic particle (such as acid) in the air, which can damage the electric wire,
the electric component and optical component surfaces. The dust in the air should be as little as
possible, or it can absorb the energy of light and be heated up. When the dust falls on the surface
of the filter, it may damage the filter. Metal particles also could damage the electric component.
Temperature: The Laser system has an optimal working temperature between 15
C~30
C.