The entrance of laser treatment room should be labeled with warning signal of laser radiation,

to prevent the entry in and exit out of the treatment room during treatment time.

Suggestion: Provide special-purpose laser operation room with air-conditioning, dehumidifier

and so on. The operator and patient have their separate regions. The caution label should be pasted
in the room before the surgical operation.

2.6 Electrical and Machine Safety

Laser treatment instrument belongs to the category 1 and type B common equipment.
When the laser treatment instrument is used by 220V or 110V, the power supply capacity is

not less than the 1000 W.

The net power source that connects with laser instrument must match international

single-phase three line AC outlets with good connection with the ground; the specification is
above 10A, connecting the ground well is very important.

When the laser treatment stops, the high-pressured component still remains high voltage.

Opening the protection cover of machine can result in electrical shock by high pressure.

Ensure the stability of the electric voltage.
Seal up all panels and the rear cover. Opening the rear cover could be dangerous as well.

2.7 Fire Prevention

The laser light device generates thermal energy. Avoid using combustible material such as

acetone or alcohol near the device. If alcohol based products are used to disinfect the laser device,
ensure that the alcohol has completely evaporated prior to operation.

Flammable and explosive objects, anesthesia, drugs, liquid, air (such as alcohol, giggle gas,

oxygen. etc.) should be avoided in the laser radiation area;

2.8 System Safety

The key switch is used to turn the unit on and off. Please remove the key when the unit is not

in use to prevent unauthorized access. The emergency cut off switch is used to shut down the
device in case of any emergency. After engaging the emergency shut off button, turn the device off
by rotating the key in a counterclockwise direction. Rotating the emergency cut off switch in the
direction indicated by the white arrows will release the button.

A 10-ampere circuit breaker is installed behind the device adjacent to the power cord. Loads

in excess of 10 amperes will trip this switch. Simply moving the switch upward will restore power
to the system. A self-diagnostic program then initializes the system and will continuously monitor
the circuitry during operation.

2.9 Transport Safety

Although the system is robust, you have to handle with care if you need to move it.
This machine laser hull inside has nice optical system. The packing should be provided by

the original factory in order to reduce the quiver in transportation, otherwise excessive shaking
may cause the maladjustment of the laser system and influence output of the energy.

Содержание QSW500

Страница 1: ...Risen Beauty Technology Co ltd Add Room619 Building 1 NO 10 Zhuoxiu North Street Fangshan District Beijing 102401 China User s Manual 1 of 26 Q Switched Nd YAG 1064 532nm Laser Treatment System QSW500...

Страница 2: ...afety 10 2 7 Fire Prevention 10 2 8 System Safety 10 2 9 Transport Safety 10 2 10 Safe Marking 11 Chapter 3 Installation 11 3 1 Equipment Detailed List 12 3 2 Installation Requirements 12 3 4 Move 15...

Страница 3: ...reet Fangshan District Beijing 102401 China User s Manual 3 of 26 6 4 Purify Soak Lip and Lips Line 25 6 5 Purify Tattoo and Wound Tattoo 25 6 6 Purify Old age Spot and Nevus 25 6 7 Purify Birthmark a...

Страница 4: ...procedure Operation Operation of the device Maintenance Maintenance information Clinical Application including application range treatment process and side effect Parameters selection Basis of paramet...

Страница 5: ...Device Brief Introduction Model name QSW500 Product name Q Switched Nd YAG Laser System Figure 1 1 shows the device Figure1 1 QSW500 Q Switched Nd YAG Laser System 1 3 Device Structure This device is...

Страница 6: ...st shell 4 Pedal control 5 Key switch 6 LCD control panel 7 Plug of Handpiece Front View of Nd YAG Laser Treatment System Notice It is recommended that only the engineer who is authorized by the manuf...

Страница 7: ...epeat frequency 1 10HZ Lead light method directly output laser The light spot diameter 2 5mm Power supply 90 130V 50Hz 60Hz or 200 260v 50Hz Environment temperature 5 40 Relative humidity 80 Cooling s...

Страница 8: ...y when not in use so as to prevent unauthorized access to the device 2 2 2 Safety Warning Only personnel trained by the distributor are authorized to maintain the inner workings of this device Any man...

Страница 9: ...he laser Note Gauze eye shield and tarsi are all effective in the protecting for the eyes 2 5 1 Laser Light Warning The Nd YAG laser is invisible infrared light Among them The laser of 532nm wavelengt...

Страница 10: ...stible material such as acetone or alcohol near the device If alcohol based products are used to disinfect the laser device ensure that the alcohol has completely evaporated prior to operation Flammab...

Страница 11: ...tion method Key switch The key switch is used shut off power This key is only provided by the manufacturer Please pull out the key after operation Please confirm that the system has been locked Emerg...

Страница 12: ...ith adequate ventilation away from devices that produce heat i e refrigerator 3 2 2 Electrical Requirements Before the Laser leaves the factory it has already been marked the local rated voltage accor...

Страница 13: ...s bag and unscrew the ventilation screw caps According to the following figure Before every treatment please ensure that there is enough water inside of the machine 4 Ensure that the key is in the off...

Страница 14: ...ater slide into the cooling system and then turn off the power 10 If the water circulation run well Then switch off the system It means you can fullyassemble system with handpiece screw on the infusio...

Страница 15: ...r the first time rinse the cooling system with distilled water 3 Ensure that the water level is adequate before treatment to prevent over heating 4 Distilled water is added periodically specially when...

Страница 16: ...the instrument and handle with care which the instrument should be supported by natural damage The markings such as Handle with care No wetness up should be on the box 3 5 2 Environment 1 Environment...

Страница 17: ...in the treatment room must wear the protective filtered glasses supplied 10 Make sure the patient is wearing the opaque protective goggles prior to treatment 11 Avoid looking directly into the intense...

Страница 18: ...18 of26 2 Treatment interface figure 5 2 Laser Energy 1 2000mJ Working frequency 1 10 Hz Return to the first interface Start to work Standby Total shooting numbers Current shooting numbers Reset the c...

Страница 19: ...ers can all be adjusted when there is something wrong with the machine when the seller give the buyer the background setting password 1 Do not point and operate the device elsewhere other than the int...

Страница 20: ...ceeds the required protection temperature the water temperature goes high it will show on the screen at the same time the buzzer will give an alarm and the instrument will cease to work When the tempe...

Страница 21: ...ity 3 The damage of system caused by fire the electric power system breaking down lightning flood disaster blast 5 1 Clean Equipment Clean the outer part of the equipment weekly with soft wet tissue Y...

Страница 22: ...head is leaks If the laser treatment head is damaged please change the laser treatment head or contact the dealer There is a cheep sound in the instrument operation process when pressing preignition T...

Страница 23: ...phage will swallow the function and the transportation function continues long tome The stabbed pigment tissue will became gradually shallow in this period time We suggest the treatment interval of 6...

Страница 24: ...e 532nm wavelength directly 5 Purify the scarlet red and brown color eyebrow can use 1064nm wavelength first then use the 532nm wavelength again After 6 weeks use the 532nm wavelength to carry on the...

Страница 25: ...as same as purify soak eyebrow 3 The bigger area only purify once time then do not purify again as far as possible Avoid the energy gathering to create the skin swelling or the blister 4 The treatmen...

Страница 26: ...y 2 Continuously strike the trouble place with the 1064nm wavelength about 15 minutes 3 The treatment time avoids June July August and September every year 4 The yellow brown spot can be cured by this...

Отзывы: