User’s Manual
10 of26
required.
The entrance of laser treatment room should be labeled with warning signal of laser radiation,
to prevent the entry in and exit out of the treatment room during treatment time.
Suggestion: Provide special-purpose laser operation room with air-conditioning, dehumidifier
and so on. The operator and patient have their separate regions. The caution label should be pasted
in the room before the surgical operation.
2.6 Electrical and Machine Safety
Laser treatment instrument belongs to the category 1 and type B common equipment.
When the laser treatment instrument is used by 220V or 110V, the power supply capacity is
not less than the 1000 W.
The net power source that connects with laser instrument must match international
single-phase three line AC outlets with good connection with the ground; the specification is
above 10A, connecting the ground well is very important.
When the laser treatment stops, the high-pressured component still remains high voltage.
Opening the protection cover of machine can result in electrical shock by high pressure.
Ensure the stability of the electric voltage.
Seal up all panels and the rear cover. Opening the rear cover could be dangerous as well.
2.7 Fire Prevention
The laser light device generates thermal energy. Avoid using combustible material such as
acetone or alcohol near the device. If alcohol based products are used to disinfect the laser device,
ensure that the alcohol has completely evaporated prior to operation.
Flammable and explosive objects, anesthesia, drugs, liquid, air (such as alcohol, giggle gas,
oxygen. etc.) should be avoided in the laser radiation area;
2.8 System Safety
The key switch is used to turn the unit on and off. Please remove the key when the unit is not
in use to prevent unauthorized access. The emergency cut off switch is used to shut down the
device in case of any emergency. After engaging the emergency shut off button, turn the device off
by rotating the key in a counterclockwise direction. Rotating the emergency cut off switch in the
direction indicated by the white arrows will release the button.
A 10-ampere circuit breaker is installed behind the device adjacent to the power cord. Loads
in excess of 10 amperes will trip this switch. Simply moving the switch upward will restore power
to the system. A self-diagnostic program then initializes the system and will continuously monitor
the circuitry during operation.
2.9 Transport Safety
Although the system is robust, you have to handle with care if you need to move it.
This machine laser hull inside has nice optical system. The packing should be provided by
the original factory in order to reduce the quiver in transportation, otherwise excessive shaking
may cause the maladjustment of the laser system and influence output of the energy.