Command
Answer
Description
S01;
Select unit 1
LDN;
0 CRLF
Start normal zero calibration
LDN?;
496,1 CRLF
Query status of the zero calibration process. “1” indicates busy.
LDN?;
496,1 CRLF
Still busy
LDN?;
396,0 CRLF
Zero calibration finished. New mV/V zero point updated.
TDD1;
0 CRLF
Save settings
Example: Calibration in mV/V
The calibration process takes some time to complete. As a result it is necessary to monitor the calibration
process to determine when it is finished. To do this, issue a LDN? query.
Command
Answer
Description
S01;
Select unit 1
VAL?;
5673 CRLF
Current reading is 0.5673 mV/V
LDN5673;
0 CRLF
Zero set to 0.5673 mV/V
LDN?;
496,1 CRLF
Query status of the zero calibration process. “1” indicates busy.
LDN?;
496,1 CRLF
Still busy
LDN?;
5673,0 CRLF
Zero calibration finished. New mV/V zero point updated.
TDD1;
0 CRLF
Save settings
21.3.27
LDW: Legacy Zero Calibration
Calibrate the zero dead weight of the scale. Due to the removal of the normal/direct calibration mode, this
command only supports normal zero calibration. For mV/V calibration see the new LDN; command (see
Section 21.3.26).
General
No. of parameters
0
Save changes
With
TDD1
Increment Trade Counter
Yes
The calibration process takes some time to complete. As a result it is necessary to monitor the calibration
process to determine when it is finished. To do this, issue a LDW? query. Refer to the calibration status table
for the LDN command in Section 21.3.26.
Example
Command
Answer
Description
S01;
Select unit 1
LDW;
0 CRLF
Start zero calibration
LDW?;
1 CRLF
Query status of the zero calibration process
LDW?;
1 CRLF
Still busy
LDW?;
0 CRLF
Zero calibration finished
TDD1;
0 CRLF
Save settings
21.3.28
LIC: Linearisation
This command gives access to the dual-point linearisation functions of the indicator. A special query LIC? is
available to verify the linearisation correction.
Important: The linearisation changes the entire scale sensitivity without zero and end point. Therefore please
double check the scale in critical load ranges.
General
139
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...