2 Maintenance Mode
SC-8000
- 4 -
<Bump Test "BUMP">
Required equipment/material
•
Bump test gas (collected in a gas sampling bag)
Connection
Connect the equipment as shown below to perform the bump test.
Gas
sampling
bag
Main unit
SC-8000
GAS IN
GAS OUT
(Open)
Sampling probe
CAUTION
The GAS OUT side of the pipe must be left open without any pipe connected. A supplied gas must
be discharged to a safe place.
WARNING
About the bump test gas
The bump test gas is a hazardous gas (toxic, oxygen deficient, etc.) therefore handle the gas and
related jigs and tools with due care. (Inhaling the gas must be avoided, the gas sampling bag must
be free of holes, etc.)
About the place for span adjustment
•
Perform span adjustment where no silicon, organic solvent, spray can gases, etc. is used.
•
Perform span adjustment indoors at normal temperatures without remarkable fluctuation (within
±5°C).
•
Perform span adjustment in an exhaust booth.