Chapter 15 Waveform Recording & Playing
RIGOL
MSO8000 User Guide
15-3
4.
Set Transition
Press
Jump To
continuously to select to jump to the recorded end frame
( Last) or the recorded start frame ( First). For settings of the start and end
frames, refer to descriptions in
.
Record Options
During the waveform recording, the oscilloscope records the waveforms of the
currently enabled channel at a specified interval until you manually stop the
recording operation or the number of recorded frames has reached the set value.
Note:
Before starting the recording operation, turn on the desired channel and
adjust the waveform display to the desired state. When starting the recording
operation, you can neither enable/disable the channel output nor adjust the
waveform display.
Before recording the waveforms, you can press
More
→
Setting
to select "Record"
to set the following parameters.
1.
Recording Interval
The recording interval indicates the time interval between the frames during the
recording process.
Press
Interval
, then rotate the multifunction knob
or use the pop-up
numeric keypad to set the time interval between frames. The available range is
from 10 ns to 10 s.
2.
Recording Frames
The recording frames refer to the number of frames that can be recorded
actually. After starting the recording operation, the oscilloscope stops the
recording operation automatically when the number of recorded frames reaches
the set value.
Press
Frames
, then rotate the multifunction knob
or use the pop-up
numeric keypad to set the number of recorded frames. The available range is
from 1 to the maximum number of frames that can be recorded currently
(available in the
Max Frames
menu). Press
Set to Max
to set the number of
recorded waveform frames to the maximum number of frames automatically.
3.
Max Frames
The menu shows the maximum number of frames that can be recorded currently.
As the capacity of the waveform memory is fixed, the more the number of points
each frame of waveform has, the less the number of waveform frames that can
be recorded. Therefore, the maximum number of recorded frames is related to
the currently selected "memory depth" (refer to
). The
current memory depth refers to the number of waveform points per frame.
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......