Chapter 5 To Trigger the Oscilloscope
RIGOL
MSO8000 User Guide
5-29
⚫
>
: triggers when the time difference (
△
T) between the specified edges of
Source A and Source B is greater than the preset time lower limit. Press
Lower
to set the delay time lower limit in Delay trigger.
⚫
<
: triggers when the time difference (
△
T) between the specified edges of
Source A and Source B is smaller than the preset time upper limit. Press
Upper
to set the delay time upper limit in Delay trigger.
⚫
<>
: triggers when the time difference (
△
T) between the specified edges of
Source A and Source B is greater than the lower limit of the preset time and
smaller than the upper limit of the preset time. Press
Upper
to set the delay
time upper limit in Delay trigger. Press
Lower
to set the delay time lower limit in
Delay trigger.
Note:
The lower time limit must be smaller than the upper time limit.
⚫
><
: triggers when the time difference (
△
T) between the specified edges of
Source A and Source B is smaller than the lower limit of the preset time or
greater than the upper limit of the preset time. Press
Upper
to set the delay
time upper limit in Delay trigger. Press
Lower
to set the delay time lower limit in
Delay trigger.
Note:
The lower time limit must be smaller than the upper time limit.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press
Mode
to quickly switch
the current trigger mode. For details, refer to descriptions in
.
Trigger Parameter Setting:
Set the trigger parameters (trigger holdoff and noise rejection) under this trigger
type. For details, refer to descriptions in
and
.
Trigger LEVEL/Threshold Level:
Rotate the
Trigger
LEVEL
knob to adjust the trigger level/threshold level. Refer
to
"Trigger LEVEL/Threshold Level"
. The current trigger level/threshold level
value is displayed at the upper-right corner of the screen.
Setup/Hold Trigger
In Setup/Hold trigger, you need to set the clock source and data source. The setup
time starts when the data signal passes the trigger level and ends at the coming of
the specified clock edge; the hold time starts at the coming of the specified clock
edge and ends when the data signal crosses the trigger level again, as shown in the
figure below. The oscilloscope triggers when the setup time or hold time is smaller
than the preset time.
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......