System
Under this menu, you can select the display Language; reset the analyzer; calibrate
the analyzer; set I/O; select reference source; set the probe power, front switch and
coupling parameters; view system information and status; set time and date as well as
manage the option license.
2.6.1.1
Language
Selects the desired display language.
2.6.1.2
Reset
Selects the instrument settings to be recalled after the analyzer is powered on. The
power on setting includes Last and Preset, and Preset has two types: Factory and
User.
1.
Power On
Sets the power on setting to Last or Preset.
When Last is selected, settings before the last power-off would be recalled
automatically when power on.
When Preset is selected, settings defined in the Preset Type would be
recalled automatically when power on.
2.
Preset Type
Sets the preset type to factory or User. The preset will be recalled after the
following operations:
Start-up the analyzer when Power On setting is “Preset”.
Press Preset
Содержание DSA1030
Страница 1: ...RIGOL User s Guide DSA1000 Series Spectrum Analyzer Sept 2014 RIGOL Technologies Inc...
Страница 2: ......
Страница 22: ......
Страница 179: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 3 Figure 3 1 Measuring a Sine signal...
Страница 181: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 5 Figure 3 2 Use the frequency counter to measure...
Страница 186: ...RIGOL Measurement Examples 3 10 User s Guide for DSA1000 Series Figure 3 4 Phase noise measurement...
Страница 188: ...RIGOL Measurement Examples 3 12 User s Guide for DSA1000 Series Figure 3 5 N dB bandwidth measurement result...
Страница 190: ...RIGOL Measurement Examples 3 14 User s Guide for DSA1000 Series Figure 3 6 Time domain power measurement...
Страница 192: ...RIGOL Measurement Examples 3 16 User s Guide for DSA1000 Series Figure 3 7 Adjacent channel power measurement result...
Страница 194: ...RIGOL Measurement Examples 3 18 User s Guide for DSA1000 Series Figure 3 8 Channel power measurement result...
Страница 196: ...RIGOL Measurement Examples 3 20 User s Guide for DSA1000 Series Figure 3 9 Occupied bandwidth measurement result...
Страница 198: ...RIGOL Measurement Examples 3 22 User s Guide for DSA1000 Series Figure 3 10 Emission bandwidth measurement result...
Страница 200: ...RIGOL Measurement Examples 3 24 User s Guide for DSA1000 Series Figure 3 11 C N Ratio measurement result...
Страница 204: ...RIGOL Measurement Examples 3 28 User s Guide for DSA1000 Series Figure 3 13 Harmonic Distortion measurement result...
Страница 211: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 35 Figure 3 17 Amplitude correction...
Страница 213: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 37 Figure 3 18 The frequency response of a filter...
Страница 214: ......