RIGOL
Chapter 2 Command Systems
Programming Guide for DSA1000A series
2-76
[:SENSe]:CORRection:CSET<n>[:STATe]?
Function
Truns on or off the specified amplitude correction.
The query returns “0” or “1”.
Explanations
<n> is 1 (Antenna), 2 (Cable), 3 (Other) or 4 (User).
Default
OFF|0
Front Panel
AMPT, Corrections, Select, Antenna/Cable/Other/User, Correction,
On/Off
13.
[:SENSe]:CORRection:CSET:TABLe:STATe
Syntax
[:SENSe]:CORRection:CSET:TABLe:STATe OFF|ON|0|1
[:SENSe]:CORRection:CSET:TABLe:STATe?
Function
Turns on or off the correction table.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
AMPT, Corrections, Corr Table, On/Off
14.
[:SENSe]:DEMod
Syntax
[:SENSe]:DEMod AM|FM|OFF
[:SENSe]:DEMod?
Function
Sets the demodulation type.
The query returns “AM”, “FM” or “OFF”.
Front Panel
Demod, Demod, AM/FM
15.
[:SENSe]:DEMod:STATe
Syntax
[:SENSe]:DEMod:STATe OFF|ON|0|1
[:SENSe]:DEMod:STATe?
Function
Turns on or off the demodulation.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Demod, Demod, AM/FM/Off
16.
[:SENSe]:DEMod:TIME
Syntax
[:SENSe]:DEMod:TIME <time>
[:SENSe]:DEMod:TIME?
Function
Sets the demodulating time, the default unit is s, such as
DEM:TIME 0.5 or DEM:TIME 500ms
The query returns “5.000000E-01”.
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...