Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-73
1.
[:SENSe]:BANDwidth:RESolution
Syntax
[:SENSe]:BANDwidth:RESolution <freq>
[:SENSe]:BANDwidth:RESolution?
Function
Sets the resolution bandwidth, the default unit is Hz, such as
:BAND:RES 1000 or :BAND:RES 1KHZ
The query returns 1000.
Explanations
<freq> ranges from 10 Hz to 1 MHz.
Default
1 MHz
Front Panel
BW, RBW
2.
[:SENSe]:BANDwidth:RESolution:AUTO
Syntax
[:SENSe]:BANDwidth:RESolution:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth:RESolution:AUTO?
Function
Enables or disables the Auto function of Resolution bandwidth.
The query returns “0” or “1”.
Explanations
In Auto mode, the resolution bandwidth decreases with the
decreasing span.
The Auto function is invalid in Zero-span mode.
Default
ON|1
Front Panel
BW, RBW, Auto/Manual
3.
[:SENSe]:BANDwidth:VIDeo
Syntax
[:SENSe]:BANDwidth:VIDeo <freq>
[:SENSe]:BANDwidth:VIDeo?
Function
Sets the video bandwidth, the default unit is Hz, such as
:BAND:VID 1000 or :BAND:VID 1KHZ
The query returns “1000”.
Explanations
<freq> ranges from 1 Hz to 3 MHz.
Default
1 MHz
Front Panel
BW, VBW
4.
[:SENSe]:BANDwidth:VIDeo:AUTO
Syntax
[:SENSe]:BANDwidth:VIDeo:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth:VIDeo:AUTO?
Function
Enables or disables the Auto function of Video bandwidth.
The query returns “0” or “1”.
Default
ON|1
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...